Aseptico AEU-20 Operation And Maintenance Manual Download Page 6

4.

5.

Torque Control

- Allows user to adjust

torque limit from maximum to minimum
levels, thus reducing the risk of instrument
breakage.  Push the Torque up/down
arrow to select the desired torque level.

Torque settings 1, 2, and 3 utilize an ‘Auto-
Reverse’ motion when torque limit is
reached to aid in freeing a stalled instru-
ment.  Once the instrument is released,
the motor resumes rotation in the previ-
ously programmed direction.

Setting the torque control to the 'MAX'
position cancels the ‘Auto-Reverse’ func-
tion and operates the motor in a continu-
ous 360° rotation at maximum torque.

Operating speed and torque level for each
ratio position are stored in the computer
memory.

For most endodontic root canal applica-
tions, a 1:8 reduction contra angle is nor-
mally used.  Proper contra angle mainte-
nance, including cleaning and lubrication,
are essential to minimize torque losses
when using the torque reduction settings.

By adjusting the torque control to reduced
limits for each instrument size, and operat-
ing at constant RPM, the risk of instrument
breakage is greatly reduced.

OPERATION FUNCTIONS - 

Cont’d

:

MAXIMUM TORQUE

Continuous 360° Rotation

HIGH TORQUE

MEDIUM TORQUE

LOW TORQUE

TORQUE

SELECTOR

AUTO-

REVERSE

ROTATION

Summary of Contents for AEU-20

Page 1: ...ENANCE INSTRUCTION MANUAL AEU 20 Endodontic System Endo ITR TM P O Box 1548 Woodinville WA 9 8333 216th Street SE Woodinville WA 98072 International 425 487 3157 Toll Free 800 426 5913 www aseptico co...

Page 2: ...ING Serious injury or death may result if ignored CAUTION Damage to property or the environ ment may result if ignored NOTE Important additional information and hints P O Box 1548 Woodinville WA 98072...

Page 3: ...ries Congratulations This system is engineered to provide many years of reliable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico...

Page 4: ...ord to the back of the console and plug into a grounded electrical receptacle 3 Connect the M 8 10K motor to the recepta cle on the lower front panel of the console 4 Attach the appropriate E Type han...

Page 5: ...ion 4 Ratio Selector Calibrates the RPM dis play to read accurately for up to four differ ent handpiece ratios Push the ratio up down arrow until the indicator light matches the handpiece you are usin...

Page 6: ...and operates the motor in a continu ous 360 rotation at maximum torque Operating speed and torque level for each ratio position are stored in the computer memory For most endodontic root canal applic...

Page 7: ...the handpiece being used i e 1 8 for a 1 8 handpiece or 1 16 for a 1 16 handpiece 3 Adjust the up down arrows on the RPM display to the desired operating speed on the RPM readout NOTE The RPM display...

Page 8: ...Pulses 3 Minimum Temperature 132 C 270 F Full Cycle Time 4 minutes Minimum Dry time 40 minutes Flash Sterilization For immediate use only C Gravity Unwrapped Minimum Temperature 132 C 270 F Full Cycl...

Page 9: ...ly with any of the above instructions may void your warranty WARNING HANDPIECES Thorough cleaning and lubrication of handpieces after each use and before sterilization is very important to ensure prop...

Page 10: ...lo fuse for 110 volts 0 16A 250V slo blo fuse for 220 volts Check motor plug connection Check foot switch connection Depress foot switch Increase RPM Check that an instrument is properly seated in the...

Page 11: ...9 NOTES...

Page 12: ...discretion to repair or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to...