6

Waste from the vacuum system will collect

in the attached plastic bottle. When the

waste container is 3/4 full it should be

emptied. After each use, the vacuum

system should be cleaned by running

clean water or a vacuum system cleanser

through the evacuator hose and

thoroughly cleaning waste bottle.

CAUTION:

Use only NON-foaming

cleansers in the vaccum lines. A small and

large capacity waste container are

available and are interchangeable. The

ADU-25 is supplied with a large capacity

container.

9. AA-25 SALIVA EJECTOR SYSTEM-

(Optional) - 

The AA-25 Saliva Ejector is

an air powered, self contained oral

evacuation system. It is supplied with a

plastic waste container and a single low

velocity hose (Fig.12).

HANDPIECE

PRESSURE

GAUGE

AA-21

CONNECTION

(Optional)

AA-25 SALIVA

EJECTOR ON/OFF

(Optional)

AIR SUPPLY

LINE

FLUSH

TOGGLE

AIR SUPPLY

PRESSURE

REGULATOR

WATER COOLANT

FLOW CONTROL

OPTIONAL

CAVITRON

CONNECTION

WATER SUPPLY TANK

(Optional)

Figure 10- Right Side of Chassis

QUICK DIS-

CONNECT

VACUUM

ON/OFF

WASTE

CONTAINER

Figure 11 - Optional AA-21 HVE

Vacuum System

OPERATION FUNCTIONS 

Cont’d

:

Summary of Contents for ADU-25

Page 1: ...OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU 25 Custom Cart...

Page 2: ...7 Handpiece Controls 4 Fig 8 Foot Control 4 Fig 9 Water Tank Removal 5 Fig 10 Right Side of Chassis 6 Fig 11 AA 21 HVE Vacuum Sys Opt 6 Fig 12 AA 25 LVE Vacuum Sys Opt 7 Fig 13 AA 290CK Central Vac K...

Page 3: ...ions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of hand held pieces or other a...

Page 4: ...E VACUUM SYSTEM Optional THREE WAY SYRINGE NWS 8 HDPE WATER SUPPLY TANK Optional HIGHSPEED HANDPIECE CONTROL AA 25 SALIVA EJECTOR SYSTEM Optional REMOVABLE INSTRUMENT TRAY 3 The ADU 25 unit must be co...

Page 5: ...connect at the rear of the delivery head Fig 4a Note manual release lever on disconnect Insert metal fitting on top of bottle into quick disconnect Locking ring will snap into place Press release leve...

Page 6: ...e locking collar surrounding the tip socket and pull the used tip straight out of the socket see Fig 6 To insert a new tip press locking collar and push tip into socket as far as it will go Release co...

Page 7: ...ol 7 FLUSH TOGGLE Fig 10 The flush toggle located on the right side of the chassis allows you to quickly and completely flush your handpieces washing away contaminants which may have accumulated in th...

Page 8: ...The ADU 25 is supplied with a large capacity container 9 AA 25 SALIVA EJECTOR SYSTEM Optional The AA 25 Saliva Ejector is an air powered self contained oral evacuation system It is supplied with a pla...

Page 9: ...clean water or vacuum system cleanser through the evacuator hose and thoroughly cleaning the waste bottle CAUTION Use only NON foaming cleansers in the vaccum lines 10 AA 290CK CENTRAL VACUUM KIT Opti...

Page 10: ...erclockwise will increase pressure 3 HANDPIECE PRESSURE ADJUSTMENT Fig 15 The amount of air pressure supplied to each handpiece may be adjusted by turning the flow control for that handpiece on the co...

Page 11: ...onds after every patient and about 20 seconds at the beginning of each day Refer to page 4 3 GENERAL CLEANING The cart exterior is designed to make cleaning and the use of prophylactic barriers as eas...

Page 12: ...rt or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service...

Reviews: