Aseptico ADU-25 Operation And Maintenance Instruction Manual Download Page 7

5

To activate the water coolant, move the

wet/dry toggle switch on the foot pedal to

the right position. Move the toggle switch

to the left position to disable water coolant.

4. WATER SUPPLY TANK (Optional) 

- The

Aseptico ADU-25 may incorporate a self

contained pressurized water system (Fig.

9). This consists of a 1-liter HDPE tank

dispensing water through the 3-way syringe

and handpiece control. The water tank is

pressurized to a constant 35 pounds per

square-inch gauge (PSIG) during use.
Before removing the tank for refilling, it

must be depressurized by turning the tank

1/4 turn counterclockwise. The tank is

removed by unscrewing it from its mount

at the bottom right rear of the chassis.

After the tank is filled, screw it back onto

the mount.  Cleaning and sanitization of

the tank is recommended between uses.

5. PRESSURE GAUGE 

(Fig. 10) - The

pressure gauge on the right panel gives

you a visual indication of the handpiece

drive pressure.

6. WATER COOLANT FLOW CONTROL

(Fig. 10) - Adjusts the flow of water coolant

to the handpiece control.

7. FLUSH TOGGLE 

(Fig. 10) - The flush

toggle located on the right side of the

chassis allows you to quickly and

completely flush your handpieces, washing

away contaminants which may have

accumulated in the handpiece and tubing.

You should flush the handpieces for about

5 seconds after every patient, and about

20 seconds at the beginning of each day

to reduce overnight bacterial accumulation

which may have occurred.
To flush your handpieces, remove them

from their holder, directing the spray away

from you and into a basin, then flip the

flush toggle and hold the desired number

of seconds.  Release the flush toggle.

8. AA-21 HIGH VOLUME VACUUM

SYSTEM (Optional)

- The AA-21 Steri-Vac

is an air powered, self contained oral

evacuator system. It is supplied with a

plastic waste container and a single high

velocity hose. (Fig. 11)
Remove the vacuum bottle and hose

assembly from packing and install into the

quick disconnect at the back of the

delivery head.
To operate, remove the vacuum hose from

holder and insert a standard oral evacuator

tip. Depress the on/off switch on the side

of the hose end for vacuum.

TURN WATER SUPPLY

TANK 1/4 TURN

COUNTERCLOCKWISE

TO RELEASE

PRESSURE BEFORE

UNSCREWING

Figure 9 - Removal of Optional Water Tank

Summary of Contents for ADU-25

Page 1: ...OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU 25 Custom Cart...

Page 2: ...7 Handpiece Controls 4 Fig 8 Foot Control 4 Fig 9 Water Tank Removal 5 Fig 10 Right Side of Chassis 6 Fig 11 AA 21 HVE Vacuum Sys Opt 6 Fig 12 AA 25 LVE Vacuum Sys Opt 7 Fig 13 AA 290CK Central Vac K...

Page 3: ...ions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of hand held pieces or other a...

Page 4: ...E VACUUM SYSTEM Optional THREE WAY SYRINGE NWS 8 HDPE WATER SUPPLY TANK Optional HIGHSPEED HANDPIECE CONTROL AA 25 SALIVA EJECTOR SYSTEM Optional REMOVABLE INSTRUMENT TRAY 3 The ADU 25 unit must be co...

Page 5: ...connect at the rear of the delivery head Fig 4a Note manual release lever on disconnect Insert metal fitting on top of bottle into quick disconnect Locking ring will snap into place Press release leve...

Page 6: ...e locking collar surrounding the tip socket and pull the used tip straight out of the socket see Fig 6 To insert a new tip press locking collar and push tip into socket as far as it will go Release co...

Page 7: ...ol 7 FLUSH TOGGLE Fig 10 The flush toggle located on the right side of the chassis allows you to quickly and completely flush your handpieces washing away contaminants which may have accumulated in th...

Page 8: ...The ADU 25 is supplied with a large capacity container 9 AA 25 SALIVA EJECTOR SYSTEM Optional The AA 25 Saliva Ejector is an air powered self contained oral evacuation system It is supplied with a pla...

Page 9: ...clean water or vacuum system cleanser through the evacuator hose and thoroughly cleaning the waste bottle CAUTION Use only NON foaming cleansers in the vaccum lines 10 AA 290CK CENTRAL VACUUM KIT Opti...

Page 10: ...erclockwise will increase pressure 3 HANDPIECE PRESSURE ADJUSTMENT Fig 15 The amount of air pressure supplied to each handpiece may be adjusted by turning the flow control for that handpiece on the co...

Page 11: ...onds after every patient and about 20 seconds at the beginning of each day Refer to page 4 3 GENERAL CLEANING The cart exterior is designed to make cleaning and the use of prophylactic barriers as eas...

Page 12: ...rt or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service...

Reviews: