Aseptico ADU-17A2 Express Air 2 Operation And Maintenance Instruction Manual Download Page 9

9.

TROUBLESHOOTING:

EXTENSION CORD RECOMMENDATIONS:

Use only a 3-wire extension cord that has a 3-blade 

grounding plug. Connect extension cord plug to a matching 

3-slot receptacle. Ensure that the extension cord is in good 

condition and that its gage wire is the correct size wire 

to carry the current this product will draw. An undersized 

cord is a potential fire hazard and will cause a drop in line 

voltage, resulting in loss of power and causing the product 

to overheat. Figure 8 indicates the correct size cord for the 

length required and the ampere rating listed on the product 

nameplate. 

If in doubt, use the next heavier gage cord. 

The smaller the gage number, the heavier the wire 

gage.

Disposing of your old product

Each Member State in the European Union (EU) requires 

that the waste from electrical and electronic products be 

disposed of separately from normal household waste. This 

includes electronics, metals and electrical accessories, 

such as signal cables or power cords. Products in this 

category carry the symbol below.  

Contact your local disposal authority and comply with the 

correct disposal method and follow national guidelines for 

unit disposal.

Unit will not start

Check system’s power connection.

Turn On/Off switch ON for operation.

Check that voltage selector module is set for source 

voltage.
Check fuse. If blown, replace with 6A/125V medium 

time delay fuse for 120V operation, or a 4A/250V slo-

blo fuse for 230V operation.

Unit starts but blows fuse

Check that voltage selector module is set for source 

voltage.
Check vacuum waste container. If full, immediately 

empty to continue operation.
Check vacuum hose assembly for blockage. 
Check that fuse is correct for source voltage.

No water coming out

Check water supply tank for water level and verify 

that cap is tight.
Depress foot switch for 5+ sec. to pressurize supply 

tank.
Ensure the handpiece air control knob is not open 

(turned clockwise) all the way.

Insufficient vacuum

Check vacuum waste container. If full, immediately 

empty to continue operation.
Check vacuum hose assembly for blockage. 
Check that waste container lid is properly seated and 

tightly secured.

Insufficient handpiece operation

Check that handpiece coolant ON/OFF toggle is 

“ON”.
Check that air control valve is turned up adequately.
Check that handpiece tubing is untangled and not 

crimped.
Check handpiece connection for missing gasket.

Compressor makes a chugging sound and does 

not start up properly

Check for sufficient power source and avoid long 

light-grade extension cords (Refer to Figure 8 Exten-

sion Cord Recommendations).

Summary of Contents for ADU-17A2 Express Air 2

Page 1: ...ADU 17A2 Express Air 2 Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...t if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60 IE...

Page 3: ...first and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments ma...

Page 4: ...he Voltage Selector Insert must be installed in the 120 position for 120V 60Hz operation or in the 240 position for 230V 50Hz operation WARNING ALWAYS DISCONNECT POWER TO THE SYSTEM BEFORE OPENING VOL...

Page 5: ...strument as shown in Figure 4 Attach the hose connector coming from the Vacuum Motor to this large white connection Connect the High Volume Vacuum HVE hose to the large gray connector located on the f...

Page 6: ...ses air Pressing both buttons simultaneously dispenses an air water mist 5 WATER SUPPLY BOTTLE The Express Air 2 incorporates a self contained pressurized water system This system consists of a 1 lite...

Page 7: ...d you have made yourself familiar with the system functions there are two operating modes to select from On Off Switch Operation Switch ON activates the vacuum and 3 way air water syringe Foot Pedal O...

Page 8: ...and between each patient use WATER LINES Disinfect the water lines weekly Prepare a 1 10 bleach solution 1 part household bleach to 9 parts water Remove water reservoir and discard residual water Rep...

Page 9: ...Check system s power connection Turn On Off switch ON for operation Check that voltage selector module is set for source voltage Check fuse If blown replace with 6A 125V medium time delay fuse for 12...

Page 10: ...Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM 1 5 Hg 39 6 liters min Water Reservoir Capacity 33 9 fl oz 1 0...

Page 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Page 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420967 Rev G ECO 15498 04 2022...

Reviews: