Aseptico ADU-17A2 Express Air 2 Operation And Maintenance Instruction Manual Download Page 8

8.

CAVITRON 

CONNECTOR

Fig. 7 - Handpiece Controls - RH

B.  Depress the Foot Pedal: The handpiece will 

activate. The Syringe, HVE, and Saliva Ejector will 

remain operational.

3.  Depress the foot pedal to operate the handpiece. Set 

the Handpiece High/Low switch to the desired position. 

For the High speed, adjust the air pressure and water 

coolant flow to the desired settings. (Turn the water 

coolant control knob 

clockwise to decrease flow

, or 

counterclockwise to increase flow

.) For the Low speed, 

adjust the air pressure to the desired setting. (See 

Figs. 6 & 7)

4.  Use the 3-Way Air/Water Syringe as necessary for 

irrigation or drying.  

5.  Adjust the vacuum pressure by moving the valves on 

both the HVE and saliva ejector vacuum heads. 

6.  Empty the Vacuum Waste Tank when it is 3/4 full.

STERILIZATION AND MAINTENANCE:

Because of its simple design, the Aseptico ADU-17A2 Express 

Air 2 requires very little maintenance. Any maintenance that is 

needed can be performed in minutes.

BLEEDING THE SYSTEM

If the unit will not be used for an extended period of time, or the 

unit might be subjected to freezing conditions, you should bleed 

the system. Simply empty the contents of the water supply tank 

and install the tank back into the cap. Operate the air/water 

syringe and handpiece with water coolant ‘ON’ until just air 

comes through the water line. Pack unit and store as normal.

HANDPIECE FLUSH

Flush the handpiece for about 5 seconds after every patient, and 

about 20 seconds at the beginning of each day.  

GENERAL CLEANING

The external surfaces of the case should be cleaned using a 

soft cloth moistened with a mild detergent solution. Any external 

surfaces of the unit that are contacted during use should be 

wiped down with a soft cloth moistened with a disinfectant at the 

beginning of each day and between each patient use.

WATER LINES

Disinfect the water lines weekly.  Prepare a 1:10 bleach solution 

(1 part household bleach to 9 parts water). Remove water 

reservoir and discard residual water. Replace empty water 

supply tank and air purge all waterlines. Fill water supply tank 

with bleach solution. Run bleach solution through all lines. Allow 

bleach solution to stand in lines for 10 minutes. Remove water 

supply tank and discard bleach. 

Flush water supply tank and 

all lines thoroughly with clean water

. Air purge and leave lines 

dry until next clinical use.

VACUUM SYSTEM

The HVE and Saliva Ejector Valves are fully autoclavable. 

Remove the valves from the hoses before autoclaving. The 

vacuum hoses should not be autoclaved. Clean hoses with 

a disinfectant solution. 

CAUTION

: Use only NON-foaming 

cleansers in the vaccum lines.

SYRINGE TIPS

The ADU-17A2 unit features a three-way air/water syringe with 

quick-change autoclavable tips. To remove a tip, press on the 

locking collar surrounding the tip socket and pull the used tip 

straight out of the socket 

(see Fig. 8). To insert a 

new tip, press locking 

collar and push tip into 

socket as far as it will 

go. Release collar and 

gently tug on tip before 

using to ensure that tip 

is securely locked into 

socket.

Syringe Tip Sterilization:

1.  Remove contaminated syringe tip.

2.  Remove all visible signs of contamination before 

autoclaving.

3.  Autoclave tip at 132° C (270° F) for ten minutes.

4.  Sterilize between each patient use.

NOTE:

 Since only the tips can be autoclaved, it is recommended 

that the Air/Water Syringe be bagged with a disposable, single-

use plastic sleeve between each patient use.

This device requires periodic maintenance to remove dust in and 

around the compressor every 12 months or if accumulations are 

seen.  The fan on the unit vents air from inside the device to the 

outside.  With the unit operating, use compressed air to spray 

into the intake vents and blow dust out of the unit.

DEPRESS LOCKING 

COLLAR

FIG. 8

Summary of Contents for ADU-17A2 Express Air 2

Page 1: ...ADU 17A2 Express Air 2 Portable Dental System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...t if ignored CAUTION Damage to property or the environment may result if ignored NOTE Important additional information and hints Conforms To AAMI STD ES60601 1 AAMI IEC STD 62366 IEC STD 80601 2 60 IE...

Page 3: ...first and all uses Clean disinfect and sterilize new or repaired instruments before first use Only use sterilized handpieces and instruments during treatment Non sterile handpieces and instruments ma...

Page 4: ...he Voltage Selector Insert must be installed in the 120 position for 120V 60Hz operation or in the 240 position for 230V 50Hz operation WARNING ALWAYS DISCONNECT POWER TO THE SYSTEM BEFORE OPENING VOL...

Page 5: ...strument as shown in Figure 4 Attach the hose connector coming from the Vacuum Motor to this large white connection Connect the High Volume Vacuum HVE hose to the large gray connector located on the f...

Page 6: ...ses air Pressing both buttons simultaneously dispenses an air water mist 5 WATER SUPPLY BOTTLE The Express Air 2 incorporates a self contained pressurized water system This system consists of a 1 lite...

Page 7: ...d you have made yourself familiar with the system functions there are two operating modes to select from On Off Switch Operation Switch ON activates the vacuum and 3 way air water syringe Foot Pedal O...

Page 8: ...and between each patient use WATER LINES Disinfect the water lines weekly Prepare a 1 10 bleach solution 1 part household bleach to 9 parts water Remove water reservoir and discard residual water Rep...

Page 9: ...Check system s power connection Turn On Off switch ON for operation Check that voltage selector module is set for source voltage Check fuse If blown replace with 6A 125V medium time delay fuse for 12...

Page 10: ...Pressure 45 50 PSI 344 kPa Average Air Flow Pressure 35 PSI High Volume Vacuum 3 0 SCFM 4 Hg 84 9 liters min Low Volume Vacuum 1 4 SCFM 1 5 Hg 39 6 liters min Water Reservoir Capacity 33 9 fl oz 1 0...

Page 11: ...r its designated and authorized warranty service center This warranty shall not apply to products 1 that have been subjected to neglect abuse misuse improper installation inadequate maintenance or dam...

Page 12: ...12 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420967 Rev G ECO 15498 04 2022...

Reviews: