Ascon tecnologic TLI40 Operating Instructions Manual Download Page 6

ASCON TECNOLOGIC - TLI 40 - OPERATING INSTRUCTIONS - 

PAG. 6

Releasing  the  key,  the  instrument  starts  normal  measurement 
acquisition once more. 

=  d.Pic: 

Pressing  the  key,  the  maximum  variation  of  the 

measurement  recorded  since  the  instrument  was  switched  on  is 
visualized on the display (highest peak  - lowest peak).   

= 0.Pot: 

For the instruments equipped with a potentiometer input, it 

is possible to set the “zero” value with this function. Pressing the key 
for  at  least  1  sec.,  the  display  will  show  the  writing 

"0.Pot"

  for 

approx.  1  sec,  and  then  "0",  assuming  the  value  measured  in  that 
instant as 0 .   

=  r.Pic: 

Pressing  the  key,  the  highest  and  lowest  peak  values  are 

re-set.  

 

= r.P0P: 

For the instruments equipped with potentiometer input, it is 

possible  to  set  the  “zero”  value  and  contemporarily  re-set  the 
highest and lowest peak values with this function . Pressing the key 
for  at  least  1  sec.,  the  display  will  show  the  writing 

"r.P0P" 

for 

approx. 1 sec., and then "0", assuming  the value measured in that 
instant as 0 and re-setting the memorized peak values.  

 

= t.Pot

: For the instruments equipped with potentiometer input, with 

this  function  it  is  possible  to  set  the  points  of  measurement  of  the 
potentiometer  by  means  of  auto-ranging  procedure  through  which 
the  parameters  “SSC",  "FSC"  and  "0.Pot."  are  automatically  re-
calculated.   
Pressing  the  key  for  at  least  1  sec.,  the  display  will  show 

"P1"

 

alternatively  to  the  value  of  the  first  point  of  setting.  Now,  position 
the potentiometer at the first point of setting and program the value 
desired for that point using the UP and DOWN keys. Once the value 
has  been  set,  press  the  P  key  :  the  instrument  will  memorize  the 
value and the display will show 

"P2"

 alternatively to the value of the 

second  point  of  setting.  Position  the  potentiometer  at  the  second 
point  of  setting  and  program  the  value  desired  for  that  point  using 
the  UP  and  DOWN  keys.  Pressing  the  P  key,  the  second  value  is 
also acquired and

 

the instrument will automatically exit from the self-

learning mode, re-calculating the measuring range. 

4.6 – DIGITAL INPUT 

The instrument can also be equipped (in alternative to OUT4) with a 
digital  input  whose  function  can  be  set  by  the  parameter 

"diF"

 

contained in the group 

]

InP”.

The parameter can be programmed as: 

= noF

 : the input does not carry out any function 

=  AaC

  :  By  closing  the  contact  connected  to  the  digital  input,  it  is 

possible to reset a memorised alarm (see par. 4.4.1) 

=  ASi: 

By  closing  the  contact  connected  to  the  digital  input,  it  is 

possible to silence an active alarm (see par. 4.4.1) 

=

 

HoLd

:  By  closing  the  contact  connected  to  the  digital  input  the 

measurement  taken  at  that  moment  is  blocked  (N.B.:  not  the 
reading  on  the  display,  therefore  the  indication  may  stabilise  itself 
with  a  delay  that  is  proportional  to  the  measuring  filter).  With  the 
hold function turned on, the instrument carries out control according 
to  the  memorised  measurement.  By  opening  the  contact,  the 
instrument starts normal measurement acquisition once more. 

=  r.Pic:

  By  closing  the  contact  connected  to  the  digital  input  the 

highest and lowest peak values are re-set.

 

= 0.Pot: 

For the instruments equipped with a potentiometer input, it 

is  possible  to  set  the  “zero”  value with this function. By closing the 
contact connected to the digital input assuming the value measured 
in that instant as 0 . 

= r.P0P: 

For the instruments equipped with potentiometer input, it is 

possible  to  set  the  “zero”  value  and  contemporarily  re-set  the 
highest  and  lowest  peak  values  with  this  function  .  By  closing  the 
contact connected to the digital input assuming  the value measured 
in that instant as 0 and re-setting the memorized peak values.

 

4.7 - RS 485 SERIAL INTERFACE 

The  instrument  can  be  equipped  with  a  RS  485  serial 
communication  interface,  by  means  of  which  it  is  possible  to 
connect  the  regulator  with  a  net  to  which  other  instruments 
(regulators  of  PLC)  are  connected,  all  depending  typically  on  a 
personal  computer  used  as  plant  supervisor.  Using  a  personal 
computer it is possible to acquire all the function information and to 
program all the instrument’s configuration parameters. The software 
protocol adopted for TLI40 is a MODBUS RTU type, widely used in 

several PLC and supervision programs available on the market (TLI 
protocol manual is available on request). 
The  interface  circuit  allows  the  connection  of  up  to  32  instruments 
on the same line. 
To  maintain  the  line  in  rest  conditions    a  120  Ohm  resistance  (Rt) 
must be connected to the end of the line. 
The instrument is equipped with two terminals called A and B which 
have  to  be  connected  with  all  the  namesake  terminals  of  the  net. 
For the wiring operation they must be interlaced with a double cable 
(telephonic  type)  and  all  the  GND  terminals  must  be  connected  to 
the ground. Nevertheless, particularly when the net results very long 
or noised and being present potential differences between the GND 
terminals, it is advisable to adopt a screened cable wired as in the 
drawing. 

If the instrument is equipped with a serial interface, the parameters 
to  be  programmed  are  the  following,  all  present  in  the  parameters 
group   

]

SEr”

 :

"Add"

  :  Address  of  the  station.  Set  a  different  number  for  each 

station, from 1 to 255. 

"baud"

 : Transmission speed (baud-rate), programmable from 1200 

to 38400 baud. All the stations have to have the same transmission 
speed. 

"PACS"

  :  Programming  access.  If  programmed  as  "LoCL"  this 

means that the instrument is only programmable from the keyboard, 
if  programmed  as  "LorE"  it  is  programmable  both  from  the 
keyboards and serial line.  
If  an  attempt  is  made to enter the programming from the keyboard 
whilst  a  communication  through  the  serial  port  is  in  progress  the 
instrument will visualise 

"buSy"

 to indicate the busy state. 

4.8 - PARAMETERS CONFIGURATION BY “A01”

 

It  is  possible  the  transfer  from  and  toward  the  instrument  of  the 
functioning  parameters  through  the  device

  A01

  with  5  poles 

connector. 
This  device  A01  it’s  mainly  useable  for  the  serial  programming  of 
the  instruments  which  need  to  have  the  same  parameters 
configuration or to keep a copy of the programming of an instrument 
and allow its rapid retransmission. 
The same device can connect the instrument via USB to a PC and 
through  the  proper  configuration  software  tools  "TECNOLOGIC 
UniversalConf”, it’s possible to configure the operating parameters. 
To  use  the  device  A01  it’s necessary that the device or instrument 
are being supplied. 

For  additional  info,  please  have  a  look  at  the  A01  instruction 
manual.

 

Summary of Contents for TLI40

Page 1: ...6 PROBLEMS MAINTENANCE AND GUARANTEE 6 1 ERROR SIGNALLING 6 2 CLEANING 6 3 GUARANTEE AND REPAIRS 7 TECHNICAL DATA 7 1 ELECTRICAL DATA 7 2 MECHANICAL DATA 7 3 MECHANICAL DIMENSIONS PANEL CUT OUT AND MO...

Page 2: ...value has been programmed push key P once more the new value will be memorised and the display will show only the code of the selected parameter By using the UP or DOWN keys it is then possible to se...

Page 3: ...ked as instrument disconnecting device which interrupts the power supply to the equipment It is also recommended that all the electrical circuits connected to the instrument must be protect properly u...

Page 4: ...ument is switched on by using the U key that has been suitably programmed see par U key with function USrb r Pic or through the suitably programmed digital input see par digital input with function dI...

Page 5: ...d in sec and the alarm will be activated only after the elapsing of that time ALARM LATCH the alarm output may behave in two different ways depending on the value added to par Abn 0 ALARM NOT LATCHED...

Page 6: ...eter input it is possible to set the zero value with this function By closing the contact connected to the digital input assuming the value measured in that instant as 0 r P0P For the instruments equi...

Page 7: ...tput 3 Par Description Range Def Note 20 O3F Functioning of output 3 ALno ALnc Ser OFF ALno Group O4 parameters relative to output 4 Par Description Range Def Note 21 O4F Functioning of output 4 ALno...

Page 8: ...to a high value when the material to be measured is particularly light reflective and reduce it when the surface is especially dark non reflective keeping in mind however that for most materials the r...

Page 9: ...ARM The alarm is activated when the process value goes below the alarm threshold set on parameter AL1L or goes higher than alarm set on parameter AL1H Ab1 ALARM AL1 FUNCTION This permits definition of...

Page 10: ...r the instruments equipped with potentiometer input with this function it is possible to set the points of measure of the potentiometer by means of an auto ranging procedure through which the paramete...

Page 11: ...ection class against electric shock Class II for Front panel Insulation Reinforced insulation between the low voltage section supply and relay outputs and the front panel Reinforced insulation between...

Page 12: ...99 99 99 1 999 9 999 0 50 mV HCFG UoLt SEnS 0 50 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 0 60 mV HCFG UoLt SEnS 0 60 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 12 60 mV HCFG UoLt SEnS 12 60...

Reviews: