Ascon tecnologic TLI40 Operating Instructions Manual Download Page 4

ASCON TECNOLOGIC - TLI 40 - OPERATING INSTRUCTIONS - 

PAG. 4

desired resolution on par. 

“dP”

  (0=1; 1=0,1; 2=0,01; 3=0,001) and 

then, on par. 

"SSC"

, the value that the instrument must visualise at 

the beginning of the scale (0/4 mA, 0/12 mV, 0/1 V o 0/2 V) and, on 
par. 

"FSC", 

the value that the instrument must visualise at the end 

of the scale (20 mA, 50 mV, 60 mV, 5 V or 10 V). 
Only  for  the  instruments  with  potentiometer  input,  the  parameters 
"SSC"  and  "FSC"  determines  the  measurement  range  and  the 
parameter   

“0.Pot”

  determines  the  zero  value  (inside  the  range 

“SSC” ... “FSC”). 
Always for the instruments with potentiometer input it is possible to 
effect the input setting through auto-ranging (see functioning of key 
“U”)  in  this  case  the  values  of  the  parameters  "SSC",  "FSC"  and 
"0.Pot" are automatically calculated by the instrument. 
In  the  case  of  infrared  sensors  (TECNOLOGIC  IRS-"A"  range),  by 
programming the sensor as "Ir.J" or "Ir.CA", the par.  "

rEFL

" is also 

present  and  it  allows  the  correction  of  possible  measuring  errors 
caused  by  the  environment  lighting  and  by  the  reflectivity  of  the 
material. This parameter should be programmed with a high value if 
the  material  to  be  measured  is  particularly  bright  /  reflective  and 
must  be  reduced  if  the  surface  is  particularly  dark  /  not  reflective, 
keeping in mind however that for most materials, the recommended 
value is within 1.00 and 0.80.   
The instrument allows for measuring calibration, which may be used 
to  recalibrate  the  instrument  according  to  application  needs,  by 
using par. 

“OFSt”

 and 

“rot”.

 

Programming  par.  “rot”=1,000,  in  par.  “OFSt”  it  is  possible  to  set  a 
positive or negative offset that is simply added to the value read by 
the  probe  before  visualisation,  which  remains  constant  for  all  the 
measurements. 
If instead, it is desired that the offset set should not be constant for 
all  the  measurements,  it  is  possible  to  operate  the  calibration  on 
any two points.  
In  this  case,  in  order  to  decide  which  values  to  program  on  par. 
“OFSt” and “rot”, the following formulae must be applied : 

“rot” = (D2-D1) / (M2-M1) “OFSt” = D2 - (“rot” x M2)

 

where: 
M1 =measured value 1 
D1 = visualisation value when the instrument measures M1 
M2 =measured value 2 
D2 = visualisation value when the instrument measures M2 
It then follows that the instrument will visualise : 

DV = MV x “rot” + “OFSt”

 

where:  DV = visualised value  

MV= measured value 

Example  1:  It  is  desired  that  the  instrument  visualises  the  value 
effectively  measured  at  20°  but  that,  at  200°,  it  visualises  a  value 
lower than 10° (190°).     
Therefore :  M1=20 ;  D1=20 ;  M2=200 ;  D2=190 
“rot” = (190 - 20) / (200 - 20) = 0,944 
“OFSt” = 190 - (0,944 x 200) = 1,2

 

Example 2: It is desired that the instrument visualises 10° whilst the 
value actually measured is 0°, but, at 500° it visualises a 50° higher 
value (550°). 
Therefore :  M1=0 ;  D1=10 ;  M2=500 ;  D2=550 
“rot” = (550 - 10) / (500 - 0) = 1,08 
“OFSt” = 550 - (1,08 x 500) = 10

 

By  using  par.

  “FiL” 

it  is  possible  to

 

program  time  constant  of  the 

software filter for the input value measured, in order to reduce noise 
sensitivity (increasing the time of reading). 
Using par. 

“diSP”, 

located in the group 

]

PAn”, 

it is possible to set

normal visualization of the display which can be the measure (dEF) 
or the alarm threshold AL1, 2, 3 or 4 (AL1, AL2, AL3 or AL4). 

4.2 - PEAK VALUES MEMORIZATION AND HOLD FUNCTION 

The  instrument  memorizes  the  highest  and  lowest  peak 
measurement values.   
To  visualize  such  values  simply  press  the 

UP

  key  to  visualize  the 

highest  peak  or  the 

DOWN

  key  for  the  lowest  peak  during  normal 

operation of the instrument.   
When the instrument is switched off, such values are always re-set. 
However, it is also possible to re-set these values if the instrument 
is  switched  on  by  using  the  U  key  that  has  been  suitably 
programmed  (see  par.    U  key  with  function  "USrb"  =  r.Pic)  or 

through the suitably programmed  digital input (see par. digital input 
with function "dIF" = r.Pic).   
Again, using the U key or the suitably programmed digital input, it is 
possible  to  visualize  the  difference  between  the  two  peaks  on  the 
display (see par. U key and digital input with function "USrb" = d.Pic 
and "dIF" = d.Pic).   
Besides  the  function  of  the  peak  values, the instrument also has a 
HOLD function through which it is possible to lock the visualization 
of the display on the measured value.   
This  function  can  be  operated  using  the  U  key  or  the  suitably 
programmed  digital  input  (see  par.    U  key  and  digital  input  with 
function "USrb" = Hold and "dIF" = Hold).   
When  the  HOLD  function  is  actived,  the  instrument  works  on  the 
alarms in operation depending on the memorized measurement. 

4.3 - OUTPUT CONFIGURATION 

The  instrument’s  outputs  (if  present)  can  be  programmed  by 
entering the group of parameters 

]

O1”, “

]

O2”, “

]

O3”, “

]

O4”

, where

different  parameters  (depending  on  the  type  of  outputs  –digital  or 
analogue- available on the instrument) are located.

 

Note:

  In  the  following  examples,  the  number  of  outputs  is 

generically indicated  with 

n

 

- DIGITAL OUTPUTS relay or SSR type :

Within the selected group only the par. 

“OnF” 

will be present

.

This parameter can be set for the following functions :
= ALno : Alarm output normally open
= ALnc : Alarm output normally closed
= OFF : Output deactivated
The coupling [outputs number – alarms number] can be effectuated
in the group relative to the alarm (“

]

AL1”,”

]

AL2”, “

]

AL3” or “

]

AL4”).

- ANALOGICAL OUTPUTS 0/4..20 mA or 0/2..10 V (only OUT1):

The parameter 

“Aor1”

 will present within the group “

]

O1”, with which

it is possible to set the beginning of the scale used for the output.
This parameter will therefore be set at:
= 0 : if one intends to use the beginning of the scale as equal to 0 (0
mA if the output is 0/4...20 mA, or 0 V if the output is  0/2...10 V)
= no_0 : if one intends to use the beginning of the scale other than 0
(4 mA if the output is 0/4...20 mA, or 2 V if the output is 0/2...10 V)
The  parameter 

“Ao1F”

  will  be  present  by  which  it  is  possible  to

configure the function of the analogical output as:
= r.inP : measurement retransmission output
= r.SEr : output led by serial communication line of the instrument
= OFF : deactivated output
In the case that the analogical output function should be configured
for  the  retransmission  of  the  signal,  it  is  therefore  necessary  to
programme  another  two  parameters  which  set  the  minimum  and
maximum reference values.
Therefore, in these cases, set the parameter 

"Ao1L"

 with the value

that the instrument must provide the minimum value (0/4 mA or 0/2
V) in output and the value to which the instrument must provide the
maximum value (20 mA o 10 V) to the parameter

 "Ao1H"

 in output.

4.4 – ALARM FUNCTION (AL1, AL2, AL3, AL4)

 

4.4.1 – ALARM OUTPUT CONFIGURATION

 

The  alarms  depend  on  the  measured  value  (AL1,  AL2,  AL3,  AL4) 
and  before  setting  them  to  work,  it  is  necessary  to  know  which 
output the alarm has to correspond to. 
First  of  all  it  is  necessary  to  configure  in  the  groups  of  parameters 

]

O  ”,

  the  parameters  relative  to  the  outputs  required  as  alarm

(“O1F” , “O2F” ,“O3F” ,“O4F”),

 

 programming the parameter relating 

to the desired output as follows : 

= ALno

 if the alarm output has to be ON when the alarm is active, 

while it is OFF when the alarm is not active 

=  ALnc

  if  the  alarm  output  has  to  be  ON  when  the  alarm  is  not 

active, while it is OFF when the alarm is active  

Note:

  In  the  following  examples  the  alarm’s  number  is  generally 

indicated as 

 n

 

Access  the  group 

]

ALn”,

  relating  to  the  alarm  to  be  set  and

program  which  output  the  alarm  signal  must  be  sent  to  on  par. 

“OALn”

The alarm functioning is instead defined by parameters : 

"ALnt "

 – ALARM TYPE 

Summary of Contents for TLI40

Page 1: ...6 PROBLEMS MAINTENANCE AND GUARANTEE 6 1 ERROR SIGNALLING 6 2 CLEANING 6 3 GUARANTEE AND REPAIRS 7 TECHNICAL DATA 7 1 ELECTRICAL DATA 7 2 MECHANICAL DATA 7 3 MECHANICAL DIMENSIONS PANEL CUT OUT AND MO...

Page 2: ...value has been programmed push key P once more the new value will be memorised and the display will show only the code of the selected parameter By using the UP or DOWN keys it is then possible to se...

Page 3: ...ked as instrument disconnecting device which interrupts the power supply to the equipment It is also recommended that all the electrical circuits connected to the instrument must be protect properly u...

Page 4: ...ument is switched on by using the U key that has been suitably programmed see par U key with function USrb r Pic or through the suitably programmed digital input see par digital input with function dI...

Page 5: ...d in sec and the alarm will be activated only after the elapsing of that time ALARM LATCH the alarm output may behave in two different ways depending on the value added to par Abn 0 ALARM NOT LATCHED...

Page 6: ...eter input it is possible to set the zero value with this function By closing the contact connected to the digital input assuming the value measured in that instant as 0 r P0P For the instruments equi...

Page 7: ...tput 3 Par Description Range Def Note 20 O3F Functioning of output 3 ALno ALnc Ser OFF ALno Group O4 parameters relative to output 4 Par Description Range Def Note 21 O4F Functioning of output 4 ALno...

Page 8: ...to a high value when the material to be measured is particularly light reflective and reduce it when the surface is especially dark non reflective keeping in mind however that for most materials the r...

Page 9: ...ARM The alarm is activated when the process value goes below the alarm threshold set on parameter AL1L or goes higher than alarm set on parameter AL1H Ab1 ALARM AL1 FUNCTION This permits definition of...

Page 10: ...r the instruments equipped with potentiometer input with this function it is possible to set the points of measure of the potentiometer by means of an auto ranging procedure through which the paramete...

Page 11: ...ection class against electric shock Class II for Front panel Insulation Reinforced insulation between the low voltage section supply and relay outputs and the front panel Reinforced insulation between...

Page 12: ...99 99 99 1 999 9 999 0 50 mV HCFG UoLt SEnS 0 50 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 0 60 mV HCFG UoLt SEnS 0 60 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 12 60 mV HCFG UoLt SEnS 12 60...

Reviews: