Appendix
1590 99 002_1 - 21. Februar 2011
83
A.3 Recommended protection distances
Recommended protection distances between portable and mobile HF telecommunications equipment and the
MeDioStar according to
IEC
60 601-1-2.
Transmitter
frequency
Equation
150 kHz to 80 MHz
d =
[
3,5
]
P
3
150 kHz to 800 MHz
d =
[
3,5
]
P
3
800 MHz to 2.5 GHz
d = [
7
]
P
3
Rated output of the
transmitter (Watt)
Protection distance
(meter)
Protection distance
(meter)
Protection distance
(meter)
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the protection distance can be estimated by using the
equation in the corresponding column, where P is the maximum output power rating of the transmitter in Watts (W)
according to the transmitter manufacturer.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.