Start up
1590 99 002_1 - 21. February 2011
21
4.2.3 General information about installation
Important information
After the installation and any relocation of the device from a cold to a warm environment
with a temperature difference of more than 5°C, allow the device to adjust to the room
temperature in unpacked condition before being used at least in the following way:
min. 2 hours at a temperature difference of up to 10°C
min. 4 hours at a temperature difference of up to 15°C
min. 8 hours at a temperature difference of more than 20°C.
Afterwards, switch the device on without a handpiece connected and allow the device to
warm up for at least 30 minutes.
Take care that the installation and operating environment of the device meets the
following conditions: temperature: 15 to 30°C, relative humidity: max. 70 % and height
not more than 2000 m above sea level.
Non-observance of these instructions may result in the destruction of the device.
For an efficient cooling of the device, please keep a minimum distance of at least 20 cm between its
lateral and rear sides to the wall.
CAUTION
Fix the wheels of the device to ensure that it cannot be moved unintentionally. Non-
observance of this instruction may cause risks of injuries.
WARNING
The laser device emits high levels of radiation in the infrared range that may cause serious
damage to the eye if the device is used improperly.
Every entrance to the laser area (usually the laser treatment room) must be equipped with
warning lamps that are lighting (or flashing) as long as the laser emits radiation to warn
persons against entering the room.
4.2.4 Door interlock connection
At the rear panel of the device, a connector is provided for the connection of a door interlock
contact to the device (see Figure 14:
).
As port a round plug hast o be used which can be screwed, type (x)V 40 according to IEC 60130-9,
e.g. Lumberg WSV 40 (as delivered).
(x): dependent on type, e.g. WS for an angled plug as delivered.
Door interlock contact 5 VDC / 20 mA
Relay contact laser emission ( max. 24 V / 0,5 A )
Figure 14: Connections for remote door interlock contact and laser warning lamp
Summary of Contents for MeDioStar
Page 1: ......
Page 2: ......
Page 3: ...User Manual MeDioStar MeDioStar NeXT...
Page 4: ...Contents 1590 99 002_1 21 Februar 2011...
Page 6: ...Contents 1590 99 002_1 21 Februar 2011...
Page 10: ......
Page 67: ...Application 1590 99 002_1 21 February 2011 57...
Page 88: ...Disposal 78 1590 99 002_1 21 February 2011 13 EC declaration of conformity...
Page 94: ......