Appendix
1590 99 002_1 - 21. Februar 2011
79
A
Appendix: EMC m
declaration
Important information
The usage of others than the following accessories, transducers and cables may result in
an increased interference emission and/or decreased immunity of the MeDioStar.
Accessory / designation
Model number
Length / dimensions
Foot switch with cable
5501 99 021
< 3.0 m
Cold device supply line
5507 04 013
< 3.0 m
Important information
This device is subject to special terms of network connections. The maximally allowed
impedance of the power grid is Z
max
= 0.1753Ohm at the power supply point. If
necessary, the local energy supplier is to be contacted to ensure that the device will be
plugged in a supply which has an impedance lower or equal Z
max
.
Important information
The MeDioStar must not be stacked or arranged directly next to or with other devices. If
the MeDioStar is used close to other devices it must be monitored to control its proper
operation.