Calibration Instructions
Please power off the scale before operating the
following procedures.
1. Turn on the scale, while showing
“111111”
“222222” “333333” “444444” “555555” ,press and
hold [
ZERO
] key; do not release your fingers until
LCD is showed
the version (6.2.XX).
2. Press [
UNIT
] key for 2 seconds to enter Calibration
Mode; LCD will be showed
SETUP
.
3. Press [
UNIT
] key to show
USUAL
, then press
again [
UNIT
] key to choose
[USUAL]
[CAL-]
with
cycle selection. Press
[NET]
key under
CAL-
to
enter calibration; LCD will be displayed the weight.
4. Input the Weight value
…
Press
[NET]
with moving flash
Press
[UNIT]
to adjust the digital value
Press
[ZERO]
after choosing to have confirmation
5. Display the AD value (dec-) after setting up the
weight value, press
[ZERO]
after the stability to
confirm the AD value for zero.
6. Put the calibration Weight and press
[ZERO]
…;
the scale can save data to the EEPROM
automatically after stability and restart this scale.