EFT/ Burst Generator PEFT Junior
4. Control elements
4.1 Generator rear panel
HAEFELY
I/O PROTECTION
SAFETY CIRCUIT
REMOTE CONTROL 730
1
2
3 4 5
6
7
LINE INPUT
POWER
8
Fig. 4.1
4.1.1 Ventilation
A single fan, mounted on the rear panel, facilitates removal of warm air from the coupling filter.
It is recommended that a space be left around the side of the PEFT Junior to assist in warm air
dissipation
4.1.2 Primary power.
Mains primary power (2) is connected to a 10A socket with integrated mains switch, fuse and
supply filter.
A mains protection fuse of the type T 0.2 A / 250 V, is mounted on the main PCB.
4.1.3 Printer connection ( PRINTER)
Test reports can be generated by connecting a serial printer to the PRINTER socket (3).
the RS232 configuration is:
Baud
rate
9600
Databit
8
Stopbit
2
Parity
0