TECPak
User’s Manual · Page 11
Name
Description
Factory
Default
Computer
Command
PID I
The integral term of the PID
loop. Will only be available if
Gain
is set to
PID
.
0.01
TEC:I
PID D
The derivative term of the PID
loop. Will only be available if
Gain
is set to
PID
.
0
TEC:D
Sensor
Sets the temperature
measurement sensor. Can be
Therm 100uA
,
Therm 10uA
,
Therm,
RTD
, or
RTD (4-wire)
.
Therm
100uA
TEC:SENS
T-Low Lim
T-Low Lim
is the lower
temperature limit. If operating in
T Mode and the actual
temperature drops below this
value, the output will be turned
off. Temperature limits not
monitored in R Mode or Ite
Mode. Limit monitoring can
also be disabled in software
(see TEC:ENAB:OUTOFF
command).
-99°C
TEC:LIM:TLO
T-High Lim
T-High Lim
is the upper
temperature limit. If operating in
T Mode and the actual
temperature rises above this
value, the output will be turned
off. Temperature limits not
monitored in R Mode or Ite
Mode. Limit monitoring can
also be disabled in software
(see TEC:ENAB:OUTOFF
command).
125°C
TEC:LIM:THI