Calibrating the Temperature Sensors

Calibration #1: Calibrating T1, T2, T3 Temperature Sensors

1

1.  Turn off the Steam QM-3 and be sure it is cold. Check that the voltage switch is in the correct position for the voltage    

 

being used.

2.  Unplug the Temperature Sensor T1 from the QM-3.

3.  Remove the Temperature Sensor T1 after unplugging it and plug it back in before install it in oil bath.

4.  Install the oil bath on stable area close to the device.

5.  Place the probe and the calibrated thermometer in the oil bath (as close as possible)

6.  Set the oil bath at 80 °C

7.  Wait that the oil bath is stabilized. And adjust its temperature to have 80 °C on the calibrated thermometer.

8.  Turn on the QM-3.

9.  On the Steam QM-3 display pad, press up and down arrows simultaneously to access the Debug sensor menu.

10.  Write the value of T1 on the calibration certificate. If T1 is +- 0,7°C regarding the set point (80°C), the sensor is 

considered as valid (go to the point 14). If not, you must adjust the sensor.

11.  On the Steam QM-3 display pad, press left and right arrows simultaneously to access the code menu.

12.  Enter 234 using the up and down arrows to change numbers and the left and right arrows to change the position to  

 

 

access temperature calibration menu, and press OK.

13.  Move the arrow to left of 80 °C on T1, press OK and wait for check mark.

14.  Write 80 °C on the adjusted column of the calibration certificate.

15.  Restart the procedure from point 5 with 140 °C instead of 80 °C.

16.  Control the calibration with random temperature between 80 °C and 140 °C (restart the procedure from point 5) and    

 

write it on the calibration certificate.

17.  If the three calibrations are correct, check the valid mark on the calibration.

1  

To reduce the time, part of calibration can be done simultaneously.

Recalibration Procedure 

->     0     0     0

                     x.xx

Code Menu

Access menu by pressing 

 and 

 

at same time.

Navigate between digits by 

pressing 

 or 

.

Change value for digit by 

pressing 

 or 

.

Press « OK ».

Turn off the steam to the Steam QM-3 and allow to cool

Version

220

1

On/Off Switch

Power cord

2

Temperature 
transmitters 

T1 and T2

  

Calibration

T Sensor

T1

80 °C

140 °C

T2

80 °C

140 °C

T3

40 °C

80 °C

Temperature Calibration Menu 234

Summary of Contents for Steam QM-3

Page 1: ...auts Sarts 2 me avenue 4 4040 HERSTAL BELGIUM T l 32 0 4 240 90 90 Fax 32 0 4 240 40 33 armstronginternational com Keep this manual with equipment for future reference STEAM QM 3 Steam Quality Monitor...

Page 2: ...ginternational com Armstrong Contents Revision History 1 Safety 2 Abbreviations and Acronyms 3 General Information 4 Recalibration Procedure 6 Calibrating the Temperature Sensors 6 Calibrating the Pre...

Page 3: ...librated Page 6 Recalibration Procedure Section Point 3 has been added Remove the Temperature Sensor T1 after unplugging it and plug it back in before install it in oil bath Calibrating the Temperatur...

Page 4: ...ualified person Installation should always be accompanied by competent technical assistance Improper installation start up operation maintenance or service may void warranty You are encouraged to cont...

Page 5: ...seconds DIN Deutsches Institut f r Normung eV dP Differential Pressure EC European Community EEC European Electrotechnical Commission EN European Norm EV Electronic Valve F Fahrenheit gal Gallon h Hou...

Page 6: ...nsors in the oil bath and recreate the actual temperatures for calibration This method is not covered in this document Pressure transmitter The pressure sensor is calibrated by simulating the voltage...

Page 7: ...ibrated on the Steam QM 3 The instruments are 3 temperature sensors 1 pressure transmitter 1 heating element 1 condensate flow meter 1 gas flow meter NCG meter 2 differential pressure sensors Heating...

Page 8: ...4 If not you must adjust the sensor 11 On the Steam QM 3 display pad press left and right arrows simultaneously to access the code menu 12 Enter 234 using the up and down arrows to change numbers and...

Page 9: ...edure in step 1 for T3 Temperature sensor NOTE Any corrections to the calibration certificate must be done following good documentation practices The error is crossed through by a single line The corr...

Page 10: ...the Steam QM 3 display pad press up and down arrows simultaneously to access the Debug sensor menu 7 Write the value of P1 on the calibration certificate If P1 is 0 1 bar regarding the set point 0 bar...

Page 11: ...d press left and right arrows simultaneously to access the code menu 7 To access the resistance calibration menu enter 069 using the up and down arrows to change numbers and the left and right arrows...

Page 12: ...alve from the draw vessel slowly to establish a cycle rate of the EV2 valve of approximately 30 seconds One cycle is from opening to closure of valve you should be able to hear this If the environment...

Page 13: ...ertificate NOTE Any corrections to the calibration certificate must be done following good documentation practices The error is crossed through by a single line The correction is written alongside wit...

Page 14: ...r syringe before resetting to 2ml 8 With the NCG injector syringe disconnected reset the syringe to 2ml 9 Connect the NCG injector syringe to the middle of the NCG injector assembly tee with a 6mm tub...

Page 15: ...w sensor calibration menu 24 On the Steam QM 3 display pad press left and right arrows simultaneously to access the code menu 25 To access the flow sensor calibration menu enter code 123 using the up...

Page 16: ...er draw vessel connecting to the bottom of the NCG burette use two wrenches 10mm and 11mm to ensure no damage to the NCG burette occurs Reconnect the expansion coil to the bottom of the NCG burette us...

Reviews: