Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 11 of 14
20. To access the flow sensor calibration menu enter code 123 using the up and down arrows to change numbers,
and the left and right arrows to change the position.
21. Move arrow to the right of Q and enter mass per cycle (using up and down arrows) then move the arrow
back to left of Q.
22. Confirm the re-calibration of the condensate flow burette by signing the relevant box on the recalibration certificate.
noTE:
Any corrections to the calibration certificate must be done following good documentation practices. The error is
crossed through by a single line. The correction is written alongside with an explanation if necessary. The correction is
signed and dated.
23. Place the 1 liter draw vessel below the EV2 valve and disconnect the 6mm tube from the inlet to the EV2 valve.
24. Reconnect the 6mm tube coming from the NCG Burette to the inlet on the EV2 valve and leave the 6mm tube
coming out of the EV2 valve in the waste container as we will use this for the calibration of the NCG Flow Meter.
Calibration
Q Sensor
-> Q
20.3
NCG
1.3
Flow sensor calibration menu 123