3
SECTION
A
ASSURE
®
PLATINUM METER SPECIFICATIONS
For additional information refer to the Assure
®
Platinum User Manual.
Product Specifications
Test strips
Assure Platinum Test Strips
Result range
20–600 mg/dL
Calibration
Plasma referenced
Sample size
Minimum 0.5 µL
Blood source
Capillary whole blood
Test time
7 seconds
Assay method
Biosensor, Glucose oxidase (
Aspergillus niger
sourced)
and Hexaammineruthenium (III) chloride
Power source
Two 1.5V alkaline AAA batteries
Battery life
More than 5,000 tests at 4 tests per day
Display type
Liquid Crystal Display (LCD)
Memory
500 results with time and date stamp
Average
7-, 14-, 30-, and 90-day averaging
Automatic shut off
2 minutes after last user action
Size
4.5” x 2.5” x 1.2”
Weight
2.8 oz. (without batteries)
Operating Ranges
Temperature
50°F–104°F (10°C–40°C)
Relative humidity
20%–80%
Hematocrit
30%–55%
Altitude
10,000 ft. (3,048 meters)
Unit of measurement
mg/dL