SECTION
B
Step 1
Insert test strip as shown. The meter will turn on automatically.
The full screen will come on for a moment. Make sure all symbols
appear on the screen. The strip symbol and flashing blood drop will
automatically be displayed.
NOTE:
If both PCS and
am
appear on the screen, perform a control
solution test.
Step 2
Press the Back
or
Forward button one time to enter the control
solution mode. A control solution bottle “
mg/dL
mmol/L
MEM
AVG
DAY
CODE
d-m-d
HYPO
am
pm
” will appear at the
top right of the screen.
NOTE:
If you do not enter the control solution mode, the control
solution result will NOT be valid.
Using the control solution mode will also flag the result in memory. This
prevents the result from being part of the averages. will prompt you to
apply control solution.
Step 3
Mix solution by gently inverting control solution bottle several times
(do not shake). Remove the cap from the control solution bottle and
place on flat surface. Squeeze the bottle and discard the first drop.
Apply the second drop to the top of the clean cap.
Step 4
Bring meter and strip to drop. Test strip will draw up the solution.
The meter will show result in 7 seconds.
Step 5
Compare the result to the range printed on the test strip bottle.
Make sure the result is within the acceptable range. If the result
falls within this range, the meter and test strip are working correctly.
NOTE:
Do not use system if control solution result is out of range.
NOTE:
Healthcare Professionals: Record result in quality logbook.
Step 6
Remove the test strip by holding the meter so that the test strip points
down. Push the strip release button forward. The test strip will fall out
and meter will shut off automatically. Dispose of the used test strip.
Wipe excess control solution from cap and replace cap on control
solution bottle.
NOTE:
Healthcare Professionals:
Repeat Steps 1–6 with a second
control solution.
Insert
contact
bars
into
meter.
Sample
Application Tip
Contact Bars
Use within 90 days (3 months) of first opening.
Usar dentro de los 90 días (3 meses) a
partir de la fecha en que fue abierto.
Control Solution Range (mg/dL)
(Rango de la solución de control)
Level 1
Level 2
Date Opened (Fecha de Apertura)
84-105
213-266
mg/dL