SECTION
D
49
Procedure:
Consult manufacturer’s direction for additional information regarding the use of the Assure
Platinum Blood Glucose Meter.
1. Wash hands.
2. Insert a test strip as shown in diagram. The meter will turn
on automatically. The full screen will come on for a moment.
Make sure all symbols appear on the screen (see page 2).
The strip symbol and flashing blood drop will automatically
be displayed.
3. Press the
or
button one time to enter the control solution
mode. You will see a control solution bottle ( )at the top right
of the screen.
Using the control solution mode will also flag the result in memory.
This prevents the result from being part of the averages.
“ ” will prompt you to apply the control solution.
4. Mix control solution by gently inverting control solution bottle
several times (do not shake). Remove the cap from the control
solution bottle. Place cap on flat surface. Squeeze the bottle and
discard the first drop. Apply the second drop to the top of the
clean cap.
5. Bring meter and strip to drop. Test strip will draw up the solution.
The meter will show result in 7 seconds.
Insert contact bars
into meter.
Sample Application Tip
Contact
Bars
NOTE:
If both PCS and
am
appear on the screen, perform a
control solution test.
NOTE: If you do not enter the control solution mode,
the control solution result will NOT be valid.