15
SECTION
B
CLEANING AND DISINFECTING THE METER
To minimize the risk of transmitting blood borne pathogens, the cleaning and disinfection
procedure should be performed as recommended in the instructions below.
The meter should be cleaned and disinfected after use on each patient.
The Assure
Prism multi Blood Glucose Monitoring System may only be used for testing multiple patients
when standard precautions and the manufacturer’s disinfection procedures are followed.
The
cleaning
procedure is needed to clean dirt, blood and other bodily fluids off the exterior of
the meter before performing the disinfection procedure.
The
disinfecting
procedure is needed to prevent the transmission of blood borne pathogens.
Only wipes with EPA registration numbers listed below have been validated for use in cleaning
and disinfecting the meter. Any disinfectant product containing these EPA registration numbers
may be used on this device. These EPA registration numbers can be found on the EPA website.
Wipes with EPA registration numbers not listed below should not be used to clean and disinfect
the Assure Prism multi Meter. Please read the manufacturer’s instructions carefully before using
their wipes on the meter.
Manufacturer
Disinfectant Brand Name
EPA #
Clorox® Professional
Products Company
Clorox Healthcare Bleach Germicidal Wipes
67619-12
Dispatch® Hospital Cleaner Disinfectant Towels
with Bleach
56392-8
Professional Disposables
International, Inc. (PDI)
Super Sani-Cloth® Germicidal Disposable Wipe 9480-4
Metrex® Research
CaviWipes1™
46781-13
Clorox and Dispatch are trademarks or registered trademarks of Clorox. Sani-cloth is a trademark or registered
trademark of PDI. Metrex and CaviWipes1 are trademarks or registered trademarks of Metrex.
Guidelines for Cleaning and Disinfecting the Assure® Prism multi Blood Glucose
Monitoring System (BGMS)
NOTE
• Do not get fluids inside the meter through the test strip port, data port or battery
compartment.
• Never immerse the meter or hold it under running water because it will damage the meter.
• Two disposable wipes will be needed for each cleaning and disinfecting procedure; one wipe
for cleaning and a second wipe for disinfecting.
Cleaning
Step 1:
Wear appropriate protective gear such as disposable gloves.
Step 2:
Open the cap of the disinfectant container and pull
out 1 towelette and close the cap.
Step 3:
Wipe the entire surface of the meter 3 times
horizontally and 3 times vertically using 1 towelette
to clean blood and other body fluids. Carefully wipe
around the test strip port by inverting the meter so
that the test strip port is facing down. This prevents
disinfectant liquid from entering the meter.
Step 4:
Properly dispose of the used towelette.
Front
Back
Both sides
• For multiple patients / professional use.
• Use with Assure Prism Multi Blood Glucose Test Strip.
• For testing outside of the body (
in vitro
diagnostic use only).
• Model: GM505RA
3V, 11 mA (CR2032 X 2)
Assure® Prism Multi Blood Glucose Meter
Made in Korea
P/N 5117-00 REV 07/14
Manufactured for
Minneapolis, MN 55439 USA
800.566.8558
www.arkrayusa.com
Contact Customer Service
at 800.566.8558, 24 hours
a day, 7 days a week, USA
and Canada.