10
Electromagnetic Compatibility
Electromagnetic Interference - Although this equipment conforms with the intent of the directive
89/336/EEC in relation to Electromagnetic Compatibility (EMC), all electrical equipment may produce
interference. If interference is suspected, move equipment away from sensitive devices or contact
the manufacturer.
Portable and mobile RF communications equipment can effect medical electrical equipment.
Radios, cell phones and similar devices may affect this equipment and should be kept at least 2 m
(6.5 ft) away from the equipment.
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information in the following tables.
The following tables document compliance levels and guidance from the IEC 60601-1-2: 2007
Standard, for the electromagnetic environment in which the
Skin IQ
MCM
should be used in a
clinical environment.
Guidance and manufacturer’s declaration - electromagnetic emissions
The
Skin IQ
MCM is intended for use in the electromagnetic environment specified below. The customer or user of the
Skin
IQ
MCM should assure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment
RF emissions
CISPR 11
Group 1
The
Skin IQ
MCM uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class B
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
yes