16
New Control Panel Icons
Power Indicator
A green light on the front panel indicates that the power
is
ON
.
Wait Indicator
If the pump detects low pressure in the mattress
replacement, the
WAIT
indicator is illuminated and the
pump performs a start up procedure to inflate the
mattress replacement.
Once correct operating pressures have been reached the
WAIT
indicator goes out.
Alarm Indicator
The pump unit incorporates a sophisticated alarm
detection system to differentiate between patient
movement and genuine alarm conditions. Whenever an
alarm occurs, the flashing red indicator is illuminated.
Additionally, an audible warning will sound unless
cancelled by the
MUTE
button as described earlier.
Once an alarm has been initiated, it can only be cancelled by switching the
pump unit off and then switching back on.
Under some circumstances the low pressure alarm may
be activated after the patient is removed from the
mattress. If this occurs the alarm may be cancelled by
switching the pump off and then switching back on.
For possible causes for the alarm, see the table in
Section 8.
Alarm Mute Button and
Indicator
An audible alarm mute is provided to cancel audible
warning sounds during an alarm condition. The yellow
light is illuminated when the alarm has been silenced.