Troubleshooting
32
001.
16
010
.33
r
e
v. 8
Problem
To check
The red “service” light is on and
flashing.
Contact your local ArjoHuntleigh representative for
maintenance.
The unit starts and stops repetitively.
If the load is greater than the unit’s safe working load,
it will not function due to an overload protection on
the motor. Batteries are low and need to be changed.
Contact your local ArjoHuntleigh representative.
The lift emits a “beep” during use. The
unit may stop lifting the patient but can
still lower the patient.
Batteries are low and need to be recharged.
Charger indicator (yellow) does not
light up when the charger is connected
to the lift.
Check that the charger is plugged into a standard
electrical outlet, and that the outlet has power. The
green indicator light on charger 700.13510.33
confirms the functionality of the charger.
The lift does not work when you press
the buttons on the hand control.
If the emergency stop is activated, gently release the
emergency button. Check if the buttons on the lift are
working. If they are working, the problem may come
from the hand control. If they are not working, check
the charge on the lift. Check to see if the hand control
is plugged in properly into the lift. The hand control
may be slightly pulled out from its socket and yet
appear as though it is plugged in. Check the
connection. If, after testing all of the above, the lift will
not operate, contact your ArjoHuntleigh
representative.