background image

4-1

4

CARE AND CLEANING

The following are ArjoHuntleigh’s recommended daily and weekly cleaning and infection control procedures for 

First Step Plus

 while in use. If this product is to be purchased, or utilized for long term rentals, it is recommended 

that manufacturer’s Infection Control Procedures be utilized by facility, dealership or home user. Contact local 
manufacturer’s representative for details.

Carefully read

 Contraindications, Risks and Precautions 

and

 Safety Tips 

sections in the 

Introduction 

chapter 

of this manual prior to performing care and cleaning procedures on 

First Step Plus

.

Follow all regulations, outlined in OSHA Occupational Exposure to Bloodborne Pathogens 29 CFR 1910.1030, 
regarding Infection Control Procedures to be used when cleaning 

First Step

 

Plus

.

DAILY CARE AND CLEANING OF 

FIRST STEP PLUS

 WHILE IN USE

NOTE:

Patient does not need to be removed from First Step

 

Plus when performing daily cleaning procedures.         

Patient may be bathed as patient support surface is being cleaned.

1.

Raise and lock Side Rails on both sides of bed, if decision has been made to use Side Rails. (Refer to 

Risks 

and Precautions

 and 

Safety Tips

 sections in the 

Introduction

 chapter of this manual.)

2.

Inflate surface fully by adjusting all three air flow control knobs to their highest settings.

3.

If Side Rails are used, lower Side Rail only on caregiver side of bed.

4.

Log roll patient to opposite side of surface.

5.

Clean patient, as required.

NOTE:

Avoid spilling fluids on Air Supply Unit.

 

Fluids remaining on controls can cause corrosion, which may 

cause components to fail or operate erratically, possibly producing potential hazards to patient and 
caregiver.

6.

Wipe Cover Sheet, as required, with soap and water or approved hospital germicide. 

7.

Move Cover Sheet for access and wipe soiled Mattress Overlay, as required. 

Replace and launder 

extensively soiled Cover Sheet or Mattress Overlay.

8.

Dry surface and Cover Sheet with towel. Remaining moisture will evaporate rapidly with air flow.

9.

Move Cover Sheet back into position.

10.

 Move patient to clean, dry side of surface and repeat Steps 4 - 11 on opposite side.

11.

 Return patient to center of surface.

12.

 Return surface to previous air pressure settings.

WEEKLY CARE AND CLEANING OF 

FIRST STEP

 

PLUS

 WHILE IN USE

In addition to daily care and cleaning described above,

 First Step Plus

 Cover Sheet should be replaced weekly or 

more often, as needed. Excessively soiled Mattress Overlay and Cover Sheet should be replaced and laundered, 
as described in 

Infection Control

 section in the 

Preparation for Use

 chapter of the 

First Step Plus

 Maintenance

 

Manual

.

Summary of Contents for First Step Plus

Page 1: ...First Step Plus OPERATIONS MANUAL P N 200676 AH Rev A 09 2013 with people in mind ...

Page 2: ...gh The electrical installation of the room must comply with the appropriate electrical wiring standards The product must be used in accordance with this manual this product s Maintenance Manual and Quick Reference Guide and all applicable labeling The only servicing to be carried out by the user is described in the Maintenance Chapter of the First Step Plus Maintenance Manual Upon customer request...

Page 3: ...p Plus 3 1 Patient Transfer from First Step Plus 3 1 Ambulatory Patient Transfer 3 1 Non Ambulatory Patient Transfer 3 1 Patient Transport 3 2 Patient Bathing 3 2 Bedpan Placement 3 3 Bedpan Removal 3 3 Repositioning Patient Using Dri Flo Underpads 3 4 Incontinence Drainage 3 4 Skin Care 3 4 General Operation 3 4 CARE AND CLEANING 4 1 Daily Care and Cleaning of First Step Plus While in Use 4 1 Wee...

Page 4: ...ii ...

Page 5: ...r death can result from the use potential entrapment or non use potential patient falls of Side Rails or other restraints See related Safety Tips section Patient Migration WARNING Specialty surfaces have different shear and support characteristics than conventional surfaces and may increase risk of patient movement sinking and or migration into hazardous positions of entrapment and or inadvertent ...

Page 6: ...ows how to get out of bed safely and if necessary how to release the Side Rails in case of fire or other emergency Monitor patients frequently to guard against patient entrapment When selecting a standard mattress ensure the distance between top of Side Rails if used and top of mattress without compression is at least 8 66 in 220 mm to help prevent inadvertent bed exit or falls Consider individual...

Page 7: ...moving parts and is not trapped under casters Improper handling of Power Cord can cause damage which may possibly produce risk of fire or electrical shock General Protocols Follow all applicable safety rules and institution protocols concerning patient and caregiver safety ...

Page 8: ...on underside of Overlay Slide elastic harness over each end of mattress Tuck tote bag under foot end of mattress Wrap elastic halter straps located at each end of Overlay under and around end of bed mattress 3 Install First Step Plus nylon fabric with moisture vapor permeable backing Cover Sheet Draw sides of Sheet down each side of bed and tuck under mattress Secure Sheet around hose connections ...

Page 9: ...ow PATIENT TRANSFER TO FIRST STEP PLUS Ambulatory Patient Transfer 1 Adjust height of both surfaces to ensure patient comfort for ingress and egress 2 Lower Side Rails when applicable 3 Lock caster wheel brakes and caster wheel steering on both surfaces 4 Assist patient in ambulatory transfer following all applicable safety rules and institution protocols 5 Center patient side to side on surface P...

Page 10: ...ght to lowest level 12 Activate Warmer and adjust temperature if desired AIR PRESSURE ADJUSTMENT 1 Adjust air flow to each section head body and feet by turning respective air flow control knob Knobs are labeled and color coded NOTE Verify air intake located on back face of Air Supply Unit is not blocked by blanket or other obstructions Air Supply Unit has thermal cutoff that will automatically de...

Page 11: ...eel steering are locked 6 Assess entrapment and fall hazards and carefully consider use of bolsters positioning aids low beds floor pads or padded Side Rail accessories when appropriate especially with confused restless or agitated patients Refer to Risks and Precautions and Safety Tips sections in the Introduction chapter of this manual 7 It is advised that capable patients are instructed on how ...

Page 12: ...manual if patient is being transferred to First Step Plus for first time PATIENT TRANSFER FROM FIRST STEP PLUS Ambulatory Patient Transfer 1 Press green Power Button to turn OFF Air Supply Unit Surface will begin to deflate 2 Adjust height of bed to same level as the surface to which patient is being transferred 3 Lock caster wheel brakes and caster wheel steering on both surfaces 4 Lower Side Rai...

Page 13: ...on caregiver side of bed CAUTION To reduce risk of patient falls during bathing caregiver must never leave patient unattended or turn away during bathing 5 Gently log roll patient toward opposite Side Rail 6 Bathe and rinse patient s exposed side and posterior NOTE Avoid spilling fluids on Air Supply Unit Fluids remaining on controls can cause corrosion which may cause components to fail or operat...

Page 14: ...ips sections in the Introduction chapter of this manual 9 Adjust head of bed for patient comfort and return bed and accessories to original configuration BEDPAN REMOVAL 1 Lower head of bed to a level position 2 Lower Side Rail on caregiver side of bed 3 Grasp bedpan firmly with one hand Using other hand gently roll patient toward opposite side of bed 4 Remove bedpan by pushing bedpan down into sur...

Page 15: ...onfigurations INCONTINENCE DRAINAGE 1 First Step Plus nylon fabric with moisture vapor permeable backing Cover Sheet should be placed over Mattress Overlay 2 Dri Flo Underpads are recommended for incontinent patients NOTE Do not use plastic backed underpads plastic will block moisture vapor transmission and air flow from Mattress Overlay Dri Flo Breathable Underpads are absorbent and do not block ...

Page 16: ...ual 2 Inflate surface fully by adjusting all three air flow control knobs to their highest settings 3 If Side Rails are used lower Side Rail only on caregiver side of bed 4 Log roll patient to opposite side of surface 5 Clean patient as required NOTE Avoid spilling fluids on Air Supply Unit Fluids remaining on controls can cause corrosion which may cause components to fail or operate erratically p...

Page 17: ...nob Knobs are labeled and color coded 2 Adjust each surface control knob until optimum pressures and patient comfort has been obtained 3 Adjust air flow so patient is cradled in a slight depression NOTE Air Supply Unit has a thermal cutoff that will automatically deactivate unit after a period of time if air intake is blocked or if unit is operated with all three air flow control knobs in OFF posi...

Page 18: ...5 in W x 5 in H Air Supply Unit 12 in L x 10 75 in H x 7 in D Weight Mattress Overlay Inflated 3 lb Air Supply Unit 11 lb Electrical Voltage 115 VAC Frequency 60 Hz Power 1 5 Amps Maximum Electrical Leakage 50 Micro Amps Power Cord 16 ft Hospital Grade Specifications subject to change without notice ...

Page 19: ...6 2 ...

Page 20: ...UET specializes in solutions therapies and products for surgical interventions and intensive care 7 QUESTIONS AND INFORMATION For questions regarding First Step Plus supplies maintenance or additional information about ArjoHuntleigh products and services please contact ArjoHuntleigh or an ArjoHuntleigh authorized representative or visit www ArjoHuntleigh com In the US call 1 800 343 0974 Originall...

Reviews: