54
Fault indications
The bed’s control software indicates problems in the electrical system by
means of flashing indicators on the Attendant Control Panel (ACP). If you
experience any of the indications below, contact ArjoHuntleigh or an
approved service agent.
Product lifetime
The lifetime of this equipment is typically ten (10) years. “Lifetime” is defined
as the period during which the product will maintain the specified
performance and safety, provided it has been maintained and operated in
conditions of normal use in accordance with the requirements in these
instructions.
Indication
Possible cause
ACP mattress platform height and head
down tilt indicators flashing
Height actuator fault (foot end)
ACP mattress platform height and foot
down tilt indicators flashing
Height actuator fault (head
end)
ACP backrest indicator flashing
Backrest actuator fault
ACP thigh section indicator flashing
Thigh section actuator fault
Mattress platform height, head down tilt,
backrest and thigh section indicators
flashing
Control unit fault