35
Function lockout
Function lockout can be used to prevent operation of the controls, e.g. when
inadvertent movement of the mattress platform could injure the patient.
To lock (prevent) or unlock (allow) functions:
Press the Function Lock button. The indicator above the button will light.
Press the ACP button(s) corresponding to the function(s) to be locked or
unlocked. The “lock” indicator LED above each function button shows its
current status:
LED on = function locked
LED off = function unlocked.
When all functions are locked or unlocked as required, press the Function
Lock button again or wait for five seconds. The indicator above the Function
Lock button will go out and the lockout settings are stored.
NOTE
When a function is locked, any associated functions are automatically disabled, e.g. locking
the backrest also disables Bio-Contour and Auto-Chair.
NOTE
Function lockout settings are retained if the bed is disconnected from the electricity supply.