21
Citadel Patient Therapy System
- Press to activate or deactivate head
body for procedures such as intubation. An audible tone will
sound to indicate when the pressure in the head section has
dropped by 50%. Periodically an audible tone will sound to
remind the user that the function is still alive.
Firm Mattress Button
to facilitate actions such as patient transfer or positioning.
An audible tone will sound when the function has completed.
Periodically an audible tone will sound to remind the user that
the function is still active. After 30 minutes the function will
cancel and the system will return to the previously selected
pressure setting.
Patient Transport Button
- Press to activate or deactivate
mattress slightly above the set pressures, prior to unplugging
the bed in preparation for transporting the patient in the bed.
Pressing the Patient Transport Button allows for continued
patient support while the system is not powered. An audible
tone will sound when the function has completed. Periodically
an audible tone will sound to remind the user that the
function is still alive.
Air Pressure Adjustment Controls
Height / Weight Presets
Press the height / weight preset button to select
the preset that most closely corresponds to the
patient body type and weight.
Pressure zone indicators will indicate pressure
settings for each preset.
Presets are provided for reference only. Individual patient
needs should be assessed and the pressure settings should be
adjusted to meet each patient’s individual needs.