8. Lift air mattress control unit and slightly tilt the air mattress control
unit forward and then gently lower it onto the frame. Carefully align
holes on top of mounting bracket with holes on frame. As air mattress
control unit sits onto the frame, gently tilt to full upright position while
control unit aligns to hole in frame. Be careful not to damage the air
connectors.
9. Gently slide air mattress control unit from side to side to align holes on
top and on front of mounting bracket with holes on extension frame.
10. Loosely install four Philips Pan Head screws through the top of
11. Install three hex socket screws through the holes on the front of
mounting bracket. These pass through the mounting bracket / frame
extension and screw into the mattress control unit. Torque screws to
10 Nm.
12. Now tighten the four Philips Pan head screws started earlier on top of
the mounting bracket. Torque screws to 10 Nm.
13.
the control unit, RH / LH lower corners.Torque screws to 10 Nm.
14.
ensure O-rings are on connection ports. Air hoses attach straight
across from the dump valve assembly over the foot extension
crossbar to the air mattress control unit in order from left to right / right
to left. A distinct click will be heard when the air connectors are fully
attached.
15.
Figure 4: Connect air hoses to mattress control unit