20
Citadel Patient Therapy System
CONTROL PANEL
min
min
min
Citadel
Patient Therapy System Control Panel
Power On / Off Button
- Press and hold for two seconds to
turn
Citadel
Patient Therapy System on or off. LED will light
and an audible tone will sound. Bed must be plugged into AC
power for control unit to turn on.
Lock Out Button
- Press and hold for two seconds to
activate or deactivate lockout of all control unit functions.
Pressing the CPR button overrides all lockouts.
Side Rail Down Indicator
- Lights up when the left or right
side rails are down.
Fowler Angle Indicator
- Lights up when fowler angle is
greater than 30 .
Alarm Indicator
- Lights up when an alarm or alarm
condition exists in the
Citadel
Patient Therapy System. The
icon will turn off when all alarm conditions are no longer
present and the alarm has been acknowledged by pressing
the Alarm Silence / Clear Button.
Alarm Silence / Clear Button
- Press to acknowledge an
alarm. Pressing this button will clear an alarm indicator for a
condition that has been resolved. If the alarm condition has
not been resolved, pressing the button will mute the alarm
for ten minutes. If the condition causing the alarm is not
corrected within ten minutes the alarm tone will resume.
- Press to activate or deactivate seat
cushions of the body section in order to assist with patient
exit and bedpan placement. An audible tone will sound to
indicate when the pressure in the seat section has dropped
by 50%. Periodically an audible tone will sound to remind the
user that the function is still active.