Before setting and activating movement detection, make sure the patient has been weighed, all additional
accessories have been accounted for using the Auto Compensation function, and the weighing display is blank.
The display will show a moving circular pattern for a few seconds then revert to showing the
patient’s weight. Press and hold either the
In Bed (1)
button or the
Egress (2)
button for two
seconds.
2QHRIWKHLQGLFDWRUVEHWZHHQWKHEXWWRQVZLOOÀDVKWRVKRZWKHVHOHFWHGPRYHPHQWGHWHFWLRQ
threshold.
Increasing movement detection threshold.
To increase the threshold, press and hold the
Egress
EXWWRQ7KHÀDVKLQJLQGLFDWRUZLOOPRYHWR
the right.
To decrease the threshold, press and hold the
In Bed
EXWWRQ7KHÀDVKLQJLQGLFDWRUZLOOPRYHWR
the left.
When the required threshold is reached, release the button.
$IWHUDIHZVHFRQGVWKHLQGLFDWRUZLOOVWRSÀDVKLQJDQGUHPDLQOLWWRVKRZWKDWPRYHPHQW
detection is activated.
,IPRYHPHQWGHWHFWLRQLVWULJJHUHGDORFDODODUPZLOOVRXQGDQGWKHWKUHVKROGLQGLFDWRUZLOOÀDVK
If the bed is connected to a suitable nurse call system, the alarm will also be indicated at the
nurse’s station.
To cancel the patient movement alarm, or to deactivate movement detection without triggering
the alarm, press the
In Bed
or
Egress
button once.
All the threshold indicators will go out to show that movement detection is deactivated.
7KHGHJUHHRIPDWWUHVVSODWIRUPSUR¿OHDGMXVWPHQWEHIRUHWULJJHULQJWKHDODUPGHSHQGVRQWKH
threshold setting.
The Auto Compensation function can be used while movement detection is activated.
The anti-entrapment system is designed to detect patient entrapment between the base and mattress platform
when the mattress platform is lowered, or is placed into tilt or Auto-Chair. The system is permanently active and
cannot be switched off.
There are four infra-red sensors, one above each castor, which generate an invisible beam around the base of the
bed.
If the beam is interrupted (e.g. by a patient’s limb) while the mattress platform is lowering, the mattress platform
stops moving and the weighing system display shows AES.
Upward movement of the mattress platform is not affected
VariZone Patient Movement Detection
Anti-Entrapment System
Before using patient movement detection, verify that the alarm can be easily heard by caregivers, e.g. at
the nurse’s station.