17
Recommended separation distances between portable
and mobile RF communications equipment and the pump
The pump is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the pump can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the pump as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d
= 1.2
√
P
80 MHz to 800 MHz
d
= 1.2
√
P
800 MHz to 2.5 GHz
d
= 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
2.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Summary of Contents for 247001
Page 2: ...This document has been translated based on 247933EN Rev 6 ...
Page 3: ...English Deutsch Español Français Ελληνικά Italiano Nederlands Svensk ...
Page 4: ......
Page 5: ...English ...
Page 7: ... ii ...
Page 28: ...Deutsch ...
Page 29: ......
Page 31: ... ii ...
Page 52: ...Español ...
Page 53: ......
Page 55: ... ii ...
Page 76: ...Français ...
Page 77: ......
Page 79: ... ii ...
Page 100: ...ΕΛΛΗΝΙΚΆ ...
Page 101: ......
Page 103: ... ii ...
Page 124: ...Italiano ...
Page 125: ......
Page 127: ... ii ...
Page 148: ...Nederlands ...
Page 149: ......
Page 151: ... ii ...
Page 172: ...Svensk ...
Page 173: ......
Page 175: ... ii ...
Page 196: ......
Page 197: ......
Page 198: ......