background image

Page | 11 

 

Commercial in confidence. Information for Use Manual Revision COU-E-REV12-080920 

4.  OPERATION 

Continued

 

 

4.8 Paper Roll Holder

 (if fitted) 

 

The Prioma couch can be equipped with a paper roll holder which (if fitted) is located at the 
head end of the couch. 

 

To Install or Change the Paper Roll: 

 

 

1.  Remove M6 hexagon nut (A), at one end of the centre rod using 10mm ring spanner. 
2.  Partly slide out centre rod, just enough to remove empty paper roll. 
3.  Insert new paper roll over centre rod. 
4.  Re-insert the rod though bracket hole and secure in position by fitting M6 hexagon nut. 
5.  Paper roll is now ready for use.  

 

 

 

5.  PRODUCT CARE

 

 

WARNING

 

Before cleaning always disconnect the couch from the mains power source. 

 

It is strongly recommended that protective clothing be used when carrying out 
any cleaning processes on the Prioma Couch. 
The following procedure also applies to the accessories. 

 

5.1 Cleaning 

Clean all surfaces with a disposable cloth soaked in a neutral detergent and hand hot 
water. Start by cleaning the upper sections of the couch working along all horizontal 
surfaces. Work methodically towards the lower sections of the couch and clean the castors 
last. Take extra care to clean areas that may trap dust or dirt. Wipe over with a new 
disposable cloth moistened with clean water and dry with disposable paper towels. 

 

5.2 Disinfecting 

After cleaning the couch as described above, wipe all surfaces with sodium 
dichloroisocyanurate (NaDCC) at a concentration of 1,000 parts per million (0.1%) of 
available chlorine. In the case of pooling body fluids, e.g. blood, the concentration of 
NaDCC should be increased to 10,000 parts per million (1%) of available chlorine. 

 

CAUTION 

Do not use abrasive compounds or pads or phenol-based disinfectant solutions. 
Do not use jet stream cleaning or wash tunnels. 

 

Summary of Contents for Prioma Couch Electric

Page 1: ...INSTRUCTIONS FOR USE COU E REV12 080920 Electric Prioma Couch Range...

Page 2: ...t 8 4 3 Backrest Adjustment 8 4 4 Knee Positioning Adjustment 8 4 5 Battery Backup if fitted 8 4 6 Central Locking Castor System 9 4 7 Fold down Arms if fitted 10 4 8 Paper Roll Holder if fitted 11 5...

Page 3: ...in the presence of flammable gases Do not use electrically operated couches in operating theatres Use the couch to support one patient only The Prioma couch is intended for one person use and may beco...

Page 4: ...hospital lines and power cables are free from possible entanglement with mechanical parts of the couch Always ensure that attendant and patient limbs are free from potential entrapment areas before op...

Page 5: ...tended Area of Use Prioma couches are appropriate for use in private practices nursing homes physio centres hospitals and clinics for the purpose of a wide range of examination and treatment procedure...

Page 6: ...ks not shown in image D 125mm Central Locking Castors x 4 E ABS Plastic Crank Arm Covers F Handset Storage Bar both sides of couch G Central Lock Pedal Neutral Brake Tracking H Upholstered Foot Sectio...

Page 7: ...e couch Disconnect the power supply cord from the electricity supply and store safely before moving the couch 3 1 Electricity Supply Connect the mains cable plug to a suitable socket outlet Make sure...

Page 8: ...tween the outer extremities of the couch and other building fixtures and equipment when operating couch functions Adjustment of the couch occurs only while the button is pressed Adjustment will stop w...

Page 9: ...t use Bed operation will be slower when the power cord is unplugged and couch is powered by the batteries CAUTION If battery backup is fitted and to ensure that the battery is kept fully charged and p...

Page 10: ...d in any direction With the pedals in the Red Down Brake position the brakes are applied to all four castors of the couch simultaneously and the couch is immobilised With the pedals in the Green Down...

Page 11: ...rioma 1 Section Arm Length 1000mm O A Fig 4 To Lower Grip and hold the Arm in the middle of the top rail A and pull up as indicated by the directional arrow fig 1 to unlock Controlling the Arm allow i...

Page 12: ...processes on the Prioma Couch The following procedure also applies to the accessories 5 1 Cleaning Clean all surfaces with a disposable cloth soaked in a neutral detergent and hand hot water Start by...

Page 13: ...the power supply plug is not damaged If either the power cable or plug is damaged then both the cable and plug must be replaced as a complete assembly by an approved service agent CASTORS 1 Apply the...

Page 14: ...s should take no more than one 1 minute to complete Visual Inspection of Control Panel Handset Casing 1 1 Inspect the outer plastic casing for any obvious cracks or damage 1 2 Inspect for damage to th...

Page 15: ...otswitch plugs disconnected Check the plugs are fully inserted in the control box Blown fuse Contact your local Arjo Service Department Damaged footswitch cable Replace footswitch Control box overheat...

Page 16: ...0 Kg 8 0 Kg With Fold down Arms Paper Roll Holder fitted excluding paper roll 13 5 Kg 9 5 Kg 9 5 Kg Safe Working Load SWL 200 Kg 200 Kg 200 Kg Maximum Patient Weight 165 Kg 165 Kg 165 Kg Electrical D...

Page 17: ...ation Label Safe Working Load Refer to Instructions for Use Date of Manufacture Product Reference Number 11 Digit Code Representing Various Options Build Types Product Serial Number Manufacturer Indic...

Page 18: ...ularly gas springs actuators batteries and other electrical devices may produce substances that are hazardous to the environment To minimise these hazards contact Arjo for information on correct dispo...

Page 19: ...regarding this product please contact your local Arjo office or approved distributor Arjo offices can be found at www arjo com int region select Have the model number and serial number of the equipme...

Page 20: ...ensure ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to...

Reviews: