5.  PRODUCT CARE 

Continued

 

 

5.3 Preventative Maintenance 

 

The following preventative maintenance procedures should be carried out every 12 
months. 
 

Other than general maintenance activities all other service or repair work must 

be performed by qualified and trained persons who are approved by Arjo.

 

 

WARNING

 

The couch must be disconnected from the mains power supply before starting any 
maintenance activity. Avoid skin contact with lubricants. 

 

GENERAL  

Examine the couch for obvious signs of damage. All aspects of the equipment 
should operate as intended. Check that all nuts, bolts and other fasteners are 
tight and are not missing. 

Examine flexible cables and conduits for cuts, cracks, abrasions or other 
deterioration. 

Check that the power supply plug is not damaged. If either the power cable or 
plug is damaged, then both the cable and plug must be replaced as a complete 
assembly by an approved service agent. 

CASTORS 

Apply the brakes and push the couch forwards. If any of the four braking 

castors rotate the brake is not fully effective.

 

Check the steering lock castor for correct operation.

 

If either the brakes or the steering locks are not working correctly contact an 

approved service agent.

 

FOLD-DOWN ARMS 

 

Inspect the Arms for general wear and tear on fittings and pivot joints every six 
months. Check for rust, structural integrity of welds and paint finish every twelve 
months. 

Raise the Arm and lock in position. Refer to the operating procedures in this 
manual for further instruction. Ensure that locking operation operates smoothly 
and check for wear and tear of pivot screw and arm posts. 

PAPER ROLL HOLDER 

Check that the paper roll holder is securely attached to the couch and centre rod 
has not been compromised in any way. 

 
 
 

Summary of Contents for Prioma Couch Electric

Page 1: ...INSTRUCTIONS FOR USE COU E REV12 080920 Electric Prioma Couch Range...

Page 2: ...t 8 4 3 Backrest Adjustment 8 4 4 Knee Positioning Adjustment 8 4 5 Battery Backup if fitted 8 4 6 Central Locking Castor System 9 4 7 Fold down Arms if fitted 10 4 8 Paper Roll Holder if fitted 11 5...

Page 3: ...in the presence of flammable gases Do not use electrically operated couches in operating theatres Use the couch to support one patient only The Prioma couch is intended for one person use and may beco...

Page 4: ...hospital lines and power cables are free from possible entanglement with mechanical parts of the couch Always ensure that attendant and patient limbs are free from potential entrapment areas before op...

Page 5: ...tended Area of Use Prioma couches are appropriate for use in private practices nursing homes physio centres hospitals and clinics for the purpose of a wide range of examination and treatment procedure...

Page 6: ...ks not shown in image D 125mm Central Locking Castors x 4 E ABS Plastic Crank Arm Covers F Handset Storage Bar both sides of couch G Central Lock Pedal Neutral Brake Tracking H Upholstered Foot Sectio...

Page 7: ...e couch Disconnect the power supply cord from the electricity supply and store safely before moving the couch 3 1 Electricity Supply Connect the mains cable plug to a suitable socket outlet Make sure...

Page 8: ...tween the outer extremities of the couch and other building fixtures and equipment when operating couch functions Adjustment of the couch occurs only while the button is pressed Adjustment will stop w...

Page 9: ...t use Bed operation will be slower when the power cord is unplugged and couch is powered by the batteries CAUTION If battery backup is fitted and to ensure that the battery is kept fully charged and p...

Page 10: ...d in any direction With the pedals in the Red Down Brake position the brakes are applied to all four castors of the couch simultaneously and the couch is immobilised With the pedals in the Green Down...

Page 11: ...rioma 1 Section Arm Length 1000mm O A Fig 4 To Lower Grip and hold the Arm in the middle of the top rail A and pull up as indicated by the directional arrow fig 1 to unlock Controlling the Arm allow i...

Page 12: ...processes on the Prioma Couch The following procedure also applies to the accessories 5 1 Cleaning Clean all surfaces with a disposable cloth soaked in a neutral detergent and hand hot water Start by...

Page 13: ...the power supply plug is not damaged If either the power cable or plug is damaged then both the cable and plug must be replaced as a complete assembly by an approved service agent CASTORS 1 Apply the...

Page 14: ...s should take no more than one 1 minute to complete Visual Inspection of Control Panel Handset Casing 1 1 Inspect the outer plastic casing for any obvious cracks or damage 1 2 Inspect for damage to th...

Page 15: ...otswitch plugs disconnected Check the plugs are fully inserted in the control box Blown fuse Contact your local Arjo Service Department Damaged footswitch cable Replace footswitch Control box overheat...

Page 16: ...0 Kg 8 0 Kg With Fold down Arms Paper Roll Holder fitted excluding paper roll 13 5 Kg 9 5 Kg 9 5 Kg Safe Working Load SWL 200 Kg 200 Kg 200 Kg Maximum Patient Weight 165 Kg 165 Kg 165 Kg Electrical D...

Page 17: ...ation Label Safe Working Load Refer to Instructions for Use Date of Manufacture Product Reference Number 11 Digit Code Representing Various Options Build Types Product Serial Number Manufacturer Indic...

Page 18: ...ularly gas springs actuators batteries and other electrical devices may produce substances that are hazardous to the environment To minimise these hazards contact Arjo for information on correct dispo...

Page 19: ...regarding this product please contact your local Arjo office or approved distributor Arjo offices can be found at www arjo com int region select Have the model number and serial number of the equipme...

Page 20: ...ensure ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to...

Reviews: