17
Electromagnetic Compliance
001.1
1720.EN rev
.13
(continued)
Guidance and Manufacturer’s Declaration -
Electromagnetic Immunity - For all Equipment and Systems
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
3 V outside ISM bands
between 0.15-80 MHz
6 V inside ISM and
amateur radio bands
between 0.15-80 MHz
3 V outside ISM bands
between 0.15-80 MHz
6 V inside ISM and
amateur radio bands
between 0.15-80 MHz
N/A
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
N/A
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz,
1 kHz sine) PM; 18 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz,
1 kHz sine) PM; 18 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
N/A