background image

How to Use the KWIKtrak Turntable

10

001.1

1720.EN rev

. 13

Automatic KWIKtrak Turntable

Operating the Turntable

1) Make sure the lift is powered.
2) Move the lift into the turntable and release the

button on the hand control. Wait for the beep

sounds.  The  turntable  will  rotate  to  the  next

position.

3) Move the lift out of the turntable along the new

path.

4) If the turntable fails to operate automatically or

in an emergency situation, manually rotate the

turntable  after  sliding  and  holding  the  quick-

release  button  in  the  open  position

(see Fig. 4). Take care for not getting injured.

Fig. 4

Manual KWIKtrak Turntable

Operating The Turntable

1) Move the lift into the center of the turntable.

2) Make  sure  that  the  lift  is  centered  with  the

turntable,  then  pull  the  Turntable’s  shortest

cord until the pivoting track is aligned with the

next position.

3) Move the lift out of the turntable along the new

path.

4) In  case  of  a  malfunction,  pull  down  the

activation arm, to unlock the pivoting track, and

rotate  the  track  manually.  Take  care  for  not

getting injured.

Adjusting Pull Cord’s Length

If the pull cords are too long, they may be 

shortened as follows.
1) Turn  the  mechanism  halfway  so  that  both

cords are of equal length.

2) Cut  both  cords  to  equal  length  making  sure

that  you  leave  at  least  20 cm  (8  in) on  each

cord.

3) Insert  one  knob  on  each  cord  and  secure  it

with a double knot.

Fig. 5

CAUTION: 

Pay  attention  to  pull  cords,  when

moving  the  ceiling  lift  onto  or  out  of  the

turntable,  to  make  sure  they  do  not  get

entangled. 

Note: 

The mechanism may block if the ceiling

lift  is  not  centered  under  the Turntable  before

pulling the cords.

Shortest possible length

͑͑͑

 20 cm

Can be cut anywhere 

within this range

Cut both cords equal

when Turntable is

turned halfway

How to Use the KWIKtrak Turntable

Summary of Contents for KWIKtrack

Page 1: ...INSTRUCTIONS FOR USE KWIKtrack Turntable 001 11720 EN rev 13 09 2019...

Page 2: ...ing to the Arjo group of companies Arjo 2019 As our policy is one of continuous improvement we reserve the right to modify designs without prior notice The content of this publication may not be copie...

Page 3: ...onmental Advice 7 Homecare Environmental Considerations 7 Product Description Labels 8 Automatic KWIKtrak Turntable 8 Manual KWIKtrak Turntable 9 How to Use the KWIKtrak Turntable 10 Automatic KWIKtra...

Page 4: ...notice any changes in the performance of your turntable or if you need any help in setting up using or maintaining your KWIKtrak Turntable The agent can offer support and service programs to maximize...

Page 5: ...on to the initial track path Product Identification The unit s identification numbers specification model serial number appear on a silver name plate fixed inside the track of the KWIKtrak Turntable H...

Page 6: ...and recycle according to the European Directive WEEE or local regulation It refers to the Instructions for Use It points out the turntable is on automatic rotation mode It points out the turntable is...

Page 7: ...pper might lead to a patient fall CAUTION Keep all components of the turntable clean and dry NOTE The Automatic KWIKtrak Turntable is designed to be used with ceiling lift series 910XXXX 911XXXX 913XX...

Page 8: ...720 EN rev 13 Automatic KWIKtrak Turntable Fig 2 1 2 3 5 4 6 Manual rotation Powered rotation Product Description Labels Legend 1 Automatic KWIKtrak Turntable 2 Quick release 3 Covers 4 Pivoting track...

Page 9: ...oduct Description Labels 001 11720 EN rev 13 Manual KWIKtrak Turntable Fig 3 1 2 3 5 4 6 Legend 1 Manual KWIKtrak Turntable 2 Activation arm 3 Activation pull cord 4 Pivoting track 5 Covers 6 End stop...

Page 10: ...the pivoting track is aligned with the next position 3 Move the lift out of the turntable along the new path 4 In case of a malfunction pull down the activation arm to unlock the pivoting track and ro...

Page 11: ...dures by Arjo and equipped with correct tools and proper documents including Parts List and Service Manual Failure to meet these requirements could result in personal injuries and or unsafe equipment...

Page 12: ...s intended after checking the previous point immediately contact your Arjo agent The turntable stops moving between track and beeps The turntable is in over current condition Verify if an object is re...

Page 13: ...er Requirements automatic model 24 VDC 1 A Designed to be powered by the lift IP rating 700 11600 IPx2 OPERATION AND STORAGE CONDITIONS Ambient temperature range Operation 5 C to 40 C 41 F to 104 F St...

Page 14: ...14 Technical Specifications 001 11720 EN rev 13 KWIKtrak Turntable Dimensions Fig 6...

Page 15: ...it could result in improper operation If such use is necessary this equipment and the other equipment should be observed to verify that they are operating normally WARNING The equipment may cause rad...

Page 16: ...rst IEC 61000 4 4 Not applicable Not applicable Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 Not applicable Not applicable Mains power quality...

Page 17: ...n 0 15 80 MHz N A Radiated RF IEC 61000 4 3 10 V m 80 MHz to 2 7 GHz 10 V m 80 MHz to 2 7 GHz N A Proximity fields from RF wireless communications equipment IEC 61000 4 3 380 390 MHz 27 V m PM 50 18 H...

Page 18: ...Page intentionally left blank...

Page 19: ...9 0 6 87426222 E mail Italy promo arjo com MIDDLE EAST Arjo Middle East FZ LLC Of ce 908 9th Floor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates D...

Page 20: ...e ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to cont...

Reviews: