HOME CARE
8
Home care
Use the Auralis system
Before using the Auralis system in home environment make sure that all caregivers, including relatives, have
read and understood the instructions in this IFU.
When the Auralis system is in use, make sure:
• The
Auralis
system is kept away from sources of heat and moisture, and protected from dust, lint and dirt.
• The
Auralis
system is not covered.
•
7KHRSHUDWLRQDOHQYLURQPHQWIXO¿OVWKHUHTXLUHPHQWVVSHFL¿HGLQVHFWLRQ³2SHUDWLQJFRQGLWLRQV´RQSDJH
52.
When the Auralis system is not in use, make sure:
• No children can access the
Auralis
system.
• No pets can come in contact with the
Auralis
system.
•
7KHVWRUDJHHQYLURQPHQWIXO¿OVWKHUHTXLUHPHQWVVSHFL¿HGLQVHFWLRQ³7UDQVSRUWDQGVWRUDJHFRQGLWLRQV´RQ
page 53.
WARNING
To avoid choking,
never leave children
unattended near the equipment. The
equipment includes small parts that may
present a choking hazard to small children,
vulnerable persons and pets if inhaled or
swallowed.
CAUTION
To avoid equipment damage, never use
the Auralis pump close to heat sources.
CAUTION
To avoid equipment damage, never use
sharp objects or electrically heated under
blankets on or under the Auralis system.
CAUTION
To avoid equipment damage, do not
expose the Auralis system, especially
WKHPDWWUHVVWRQDNHGÀDPHVVXFKDV
cigarettes. A leak in the Auralis mattress or
Auralis seat cushion could propagate the
¿UH
CAUTION
To avoid equipment damage, pets and
children must be supervised in the vicinity
of the Auralis system.
WARNING
Do not attempt to service or maintain the
Auralis pump while it is in use.
WARNING
Electrical equipment may be hazardous
if misused. Do not use unapproved
accessories or attempt to modify,
disassemble or otherwise misuse the
Auralis system
WARNING
To avoid discomfor t or injur y for the
patient, ALWAYS set the comfor t level
at maximum for patients weighing over
360 kg (793 lb).