TROUBLESHOOTING AND ALARMS
49
Continue on the next page.
INDICATOR
POSSIBLE CAUSE
SOLUTION
ALARM
ACTIVATION
DELAY
POWER
FAILURE
Power source failure.
Check if power is available and
check the power cable.
When the battery is fully
charged the Auralis
pump works
for a minimum of 3 hours.
Less than 2 seconds
BATTERY LOW
Less than 1 hour left on the
battery.
Charge the battery by using
mains power supply to run the
Auralis
pump.
500 ms
READY FOR
SERVICE
and
Front panel
alarms show
green
The
Auralis pump needs
service.
&RQWDFWTXDOL¿HGVHUYLFH
personnel, quoting the serial
number.
7R¿QGWKHVHULDOQXPEHUVIRU
the Auralis
pump, mattress
or seat cushion see sections
³/DEHOV´RQSDJHDQG
³/DEHOVRQ$XUDOLVPDWWUHVVHV
Auralis seat cushion and Auralis
PDWWUHVVH[WHQVLRQV´RQSDJH
58
.
Immediate upon
expiry of service timer
INTERNAL
FAULT
and
Front panel
alarms show
yellow
The
Auralis pump has
detected an internal fault.
Switch the Auralis pump off
and restart it. If alarm persists,
FRQWDFWTXDOL¿HGVHUYLFH
personnel.
Maximum 22 minutes
AUTOFIRM
does not
activate
1. The Auralis seat cushion
is in use.
2. The Auralis mattress is in
use.
1. Disconnect the Auralis seat
cushion.
7KH$XWR¿UPPRGH
only applies to the Auralis
mattresses.
2.
7RDFWLYDWH$XWR¿UPPRGH
SUHVVDQGKROGWKH$XWR¿UP
button for 2 seconds.
Not applicable