16
14.10.2003
14.148.- / A414D en / Software 1.30
A sample of a monitoring session:
- Switch the pump on with an inserted filled infusion set.
- Type "Ctrl+N" to set the pump in remote mode.
- Type "Ctrl+B", then "01200200" and then "Ctrl+C" which sets the rate to 12.0 ml/h and an infusion total of
200 ml. The rate should now be shown in the left display of the pump.
- Type "Ctrl+R" to start the infusion.
- Type "Ctrl+T" to stop the infusion.
Format statusbyte-2:
Reserved
Parity
Always high
Occlusion, pump stop
Reserved
Bottle empty
, pump stop
T
oo many drops, pump stop
P 1 X X X X X X
Airbubble, pump stop
Format statusbyte-3:
Global Alarm
Parity
Always high
Remote mode active
Door open
Pump stop (KVO)
Reserved
P 1 X X X X X X
KVO active
Format statusbyte-4:
Standby alarm active
Data lock active
Parity
Always high
Reserved
To
tal volume reached, pump stop (KVO)
Ti
mer alarm, pump stop (KVO)
Bolus active
P 1 X X X X X X
Format statusbyte-5:
P 1 X X X X X X
Format statusbyte-1:
Battery active
Always high
Pump type (0 =
A414)
Battery low prealarm
Battery low
, pump stop
Reserved
P 1 X X X X X X
Battery defective
Bargraph-LED lower+2
Bargraph-LED lower+3
Parity
Always high
Pressure indicator (adress 46)
Bargraph-LED lower+1
Bargraph-LED upper
Bargraph-LED lower
Parity