14.148.- / A414D en / Software 1.30
14.10.2003
23
5
Safety standard check
Safety standard check
ARGUS 200
O
ARGUS Medical AG
ARGUS 400
O
ARGUS 404
O
Serial-no:
........................
ARGUS 414
O
Hospital/Dept./Customer:
The safety standard check has to be performed at least every 24 months or after 10000 hours of operation.
The check has to be done in accordance to the operation- and service manuals.
1
Visual check for damage, cleanness and completeness:
- Housing, labels, accessories, connectors,
power cable, etc.
2
Test the function of the stop flow lever:
3
Keep key "START" (ARGUS 200/400/404) or
- Display shows the software release:
V
..................
"MODE" (ARGUS 414) pressed while switching on
- Display of 2, 4, 7, F., ml total, ml inf., h.min
the pump.
- Test of the red alarm LEDs:
Pressure display, air, battery, drops, ALARM
- Test of the green operation LEDs:
Line, battery, drops, KVO
4
Place a filled tube in the air detector:
- LED air alarm disappeared
5
Set rate to 111 ml/h, press "START/STOP"
- After 12 sec. the acoustical alarm and
(without drop detector):
LED drop-alarm + LED ALARM released
6
Press key "MODE":
- Acoustical alarm switches off
7
Simulate drops manually:
- LED drops (green) lights up
8
Check of the occlusion-alarm-pressure:
- See Service manual "Replacement of parts"
Infusion set:
Codan .....
Braun .....
Other ...........................
Pressure increase to ≥1.1 bar?
Test of the occlusion-alarm-pressure
Preset level: ......... mbar
Measured: ......... mbar
9
Check of the pump accuracy:
- See service manual "Replacement of parts"
Rate: 250 ml/h
Preset volume: 25 ml
Measured volume: ............ ml
10 Test the pump at max. rate (999 ml/h):
- Running smooth?
11 Battery check by setting the rate to 60 ml/h,
- Green LED battery light?
disconnect the line and start the pump:
- Battery prealarm after typ. 4 hrs?
(Red LED battery alarm + acoustical alarm)
Running time: ............. min
- Battery alarm 15 min. after prealarm
(If the specified typical 4hrs of operation are not
(Red LED battery alarm + ALARM + alarm acoustically)
required, the battery has to be changed only
- After 6 min. the pump switches off
if the time is <3 hrs, due to environmental pollution)
12 Charge the battery min. 16 hrs.
13 Check the external connections:
- Staff alerting system
- External drop detector
- Computer interface RS232
14 Electrical test according to EN60601-1
- Measurements attached
(all measurements made with a power cable 2,5 m)
The pump has passed the safety standard check and is safe for use.
Date:
...........................
Name:
...............................
Signature:
....................................
................................................................................................................................................................................................
...................................................................................................................................................