29
Advanced settings
Enter advanced settings
Press “ENTER” to access menu
- Press until “advanced
settings” appears in the display
- Press “ENTER!
Press “ENTER” to type password
Factory settings for the password
is “0000”
Press “Esc” to exit menu/go back.
The display will return to the main screen after 1 minute of inactivity.
When you have entered the Advanced settings successfully, you will have the
following options:
Calibration
Temperature probes are mounted inside the appliance. Probe 1 controls the
cooling system and monitors the temperature inside the appliance.
To ensure precise and consistent operation, the probes can be calibrated to
offset potential changes and drifts in measurements. We recommend you to
calibrate the probe temperatures once a year.
Automatic defrost
Automatic defrost is
NOT
possible on the ULUF range. – Please see
Maintenance /defrosting.
Manual defrost
The manual defrost function of the G214 controller is
NOT
possible to be
enabled on the ULUF range. – Please see Maintenance /defrosting.
Probe at main screen
The temperature probe of the appliance (Probe 1) controls the cooling system.
It is possible to add up to two optional extra probes (Probe 2 and Probe 3) for
temperature monitoring.
The chamber probe is measuring the air temperature inside the appliance to
control the operation of the cooling system, when the compressor turns on
and off the temperature will rise and fall accordingly.
To get a better monitoring of the actual temperature of the products stored
in the appliance, probe 2 and/or probe 3 can be installed and placed in a
glycerol-filled bottle inside the appliance.
The probe related to the temperature monitoring shown on the main screen
is changed in this menu. The probe, which is selected, is shown above the
temperature in the main screen.
The temperature alarms will react to the probe selected in this menu. All three
probes can be calibrated in the “Calibration” menu. The probe displayed on
the main screen can be changed in this menu.
Attention!
Only Probe 1 is used for temperature control. Do not move or exchange for one of the
other probes as this may affect the appliance’s performance.
- Press until the “Probe
at Main Screen” appears in the
display
- Press “ENTER” to access
- Use change the value
- Use to highlight “save”
Press “ENTER” to save
Press “ENTER” to access menu
Press “ENTER” to access
custom settings
Press “ENTER” to type password
The password is “0000” -
Press “ENTER” to enter the
highlighted number and press
“ENTER” to continue