USER MANUAL

MINI CENTRIFUGE

1. Safety

Follow precautions and all the safety requirements described on this user 

manual to prevent any damages and failure of equipment.

1-1. Caution: Installation 

1. The device must be kept in a dry and well-ventilated room.

2. According to IEC61010-2-020 maintain a 30cm “clearance envelope” around the device 

    while the rotor is spinning. 

3. Do not operate the device in areas where work is completed with explosice substances

4. Connect the device to voltage sources which correspond to the electrical requirements

    on the label attached to the device.

5. Do not install the device at a place with high voltage regulation.  

    (Allowable voltage regulation limits  : ± 10 %) 

6. Use sockets with a protective earth conductor and suitable power cord.

Summary of Contents for Smart 6

Page 1: ...ARA Smart 6...

Page 2: ...ARA Smart 6 Research Use Only Copyright 2019 Hanil Scientific Inc All rights reserved Contact Us If you have any questions contact Hanil Scientific Inc techsupport ihanil com info ihanil com...

Page 3: ...tended Use 5 3 Product overview 6 4 Technical specifications 7 5 Installaion Operation 8 5 1 Power connection 8 5 2 Operation 9 6 Troubleshooting 10 7 Rotor 11 CONTENTS The appearance or specification...

Page 4: ...2 According to IEC61010 2 020 maintain a 30cm clearance envelope around the device while the rotor is spinning 3 Do not operate the device in areas where work is completed with explosice substances 4...

Page 5: ...nd contact Hanil Technical Support Team 8 Do not block vents 9 Do not drop or subject the device to severe impacts 10 Caution when handling the AC DC adaptor Do not bend or pull the AC DC adaptor Use...

Page 6: ...pen button Power Connector Running Mode Selection Switch Operation mode switch Lid Closing the lid Start the centrifugation Opening the lid Stop the centrifugation Short Mode S Stop the centrifugation...

Page 7: ...Rotor 1 5 2 0 mL x 6 hole Rotor 2 PCR Strips 0 2 mL 0 5 mL Adaptor Optional Max RPM 6 500 RPM Max RCF Angle Rotor 1 5 2 0 mL x 6 hole 2 234 xg 2 PCR Strips Rotor 1 412 xg Radius Angle Rotor 1 5 2 0 mL...

Page 8: ...adaptor Keep this manual near the device at all times 1 Connect the power cable to the power socket on the back side of the device and plug the cord into the outlet 2 A beep sounds when the power is...

Page 9: ...s the lid open button 3 Opening the lid 2 Starting Stopping the centrifugation Short Mode S When the lid is closed centrifutaion starts Operataion stops after 12 sec of operation Continuous Mode C Ope...

Page 10: ...ed 1 Power failure Check the mains connection 2 AC adaptor defect Check the output DC volt of the adapter using a multimeter 3 Lid sensor failure Contact Hanil Technical Support Team 4 If the beep doe...

Page 11: ...A2 0 6 Hole angle 40 Max Capacity 6 x 1 5 2 0 mL Size x H 88 x 24 5 mm 1 5 2 0 11 47 3 6 500 2 234 0 5 TR0 5 2 8 x 31 36 2 6 500 1 710 0 2 TR0 2 2 6 5 x 23 29 2 6 500 1 379 A0 2 16 Hole angle 40 Max...