ARA Smart 6 Manual Download Page 11

7. Rotor

Rotor

Tube 
Capacity
(mL)

Required
Adaptor

Bore Ø
Radius 
(mm)

Max. 
RPM(rpm)
Max. RCF (xg)

  A2.0-6

Hole angle : 
∠40°
Max. Capacity : 
6 x 1.5/2.0 mL
Size (ø x H) : 
88 x 24.5 mm

1.5

/2.0

-

11

47.3

6,500
2,234

0.5

TR0.5(2)

8 x 31

36.2

6,500
1,710

0.2

TR0.2(2)

6.5 x 23

29.2

6,500
1,379

  A0.2-16

Hole angle : 
∠40°
Max. Capacity : 
16 x 0.2 mL
Size (W x D x H) : 
73.5 x 55.5 x 21

0.2 mL

(PCR 

rotor)

-

6.2

6,500
2,182

Summary of Contents for Smart 6

Page 1: ...ARA Smart 6...

Page 2: ...ARA Smart 6 Research Use Only Copyright 2019 Hanil Scientific Inc All rights reserved Contact Us If you have any questions contact Hanil Scientific Inc techsupport ihanil com info ihanil com...

Page 3: ...tended Use 5 3 Product overview 6 4 Technical specifications 7 5 Installaion Operation 8 5 1 Power connection 8 5 2 Operation 9 6 Troubleshooting 10 7 Rotor 11 CONTENTS The appearance or specification...

Page 4: ...2 According to IEC61010 2 020 maintain a 30cm clearance envelope around the device while the rotor is spinning 3 Do not operate the device in areas where work is completed with explosice substances 4...

Page 5: ...nd contact Hanil Technical Support Team 8 Do not block vents 9 Do not drop or subject the device to severe impacts 10 Caution when handling the AC DC adaptor Do not bend or pull the AC DC adaptor Use...

Page 6: ...pen button Power Connector Running Mode Selection Switch Operation mode switch Lid Closing the lid Start the centrifugation Opening the lid Stop the centrifugation Short Mode S Stop the centrifugation...

Page 7: ...Rotor 1 5 2 0 mL x 6 hole Rotor 2 PCR Strips 0 2 mL 0 5 mL Adaptor Optional Max RPM 6 500 RPM Max RCF Angle Rotor 1 5 2 0 mL x 6 hole 2 234 xg 2 PCR Strips Rotor 1 412 xg Radius Angle Rotor 1 5 2 0 mL...

Page 8: ...adaptor Keep this manual near the device at all times 1 Connect the power cable to the power socket on the back side of the device and plug the cord into the outlet 2 A beep sounds when the power is...

Page 9: ...s the lid open button 3 Opening the lid 2 Starting Stopping the centrifugation Short Mode S When the lid is closed centrifutaion starts Operataion stops after 12 sec of operation Continuous Mode C Ope...

Page 10: ...ed 1 Power failure Check the mains connection 2 AC adaptor defect Check the output DC volt of the adapter using a multimeter 3 Lid sensor failure Contact Hanil Technical Support Team 4 If the beep doe...

Page 11: ...A2 0 6 Hole angle 40 Max Capacity 6 x 1 5 2 0 mL Size x H 88 x 24 5 mm 1 5 2 0 11 47 3 6 500 2 234 0 5 TR0 5 2 8 x 31 36 2 6 500 1 710 0 2 TR0 2 2 6 5 x 23 29 2 6 500 1 379 A0 2 16 Hole angle 40 Max...