ADSP7000634
rev. 1.3
23/09/2020
17/36
3.3.4
S
ETTING THE
T
EMPERATURE
The temperature setting allows you to correct the conductivity value read.
From the Programming Menu, after the entry
InP
, by accessing the entry
°C
, it is possible to choose between
manual setting "
noA
" and automatic temperature reading "
Aut
" from the
PT100 probe.
If you want to correct the measured value in relation to the thermal variation,
in manual mode, just set a temperature equal to 25°C.
The default setting is equivalent to "
noA
", with temperature equal to 25°C.
From the entry "
°C
", by pressing
Enter
, it is possible to display:
Temperature value measured from 0.0 to 99.9°C if the mode is
Aut
:
if the probe is not connected, the display will show “
t - -
“;
Set temperature value, from 0 to 99.9 °C if the mode is
noA;
To go back to the Programming Menu, press
Enter
again.
F
IG
.
15-
T
EMPERATURE
3.3.5
C
ALIBRATING THE
P
UMP
Pump calibration allows you to achieve more accurate dosing, as it allows you to correct the dosing speed
calculation in relation to the actual flow rate of the pump. If calibration has never been carried out, the dosing speed
percentage is calculated based on the pump rated flow rate value (read the value on the label).
From the
tAr
entry, you will be able to calibrate the pump for a fixed time of 60 seconds, thereby measuring the
amount of liquid dosed via a graduated container used as a reference: from the start label “
StA
”, launch the count
by pressing
Enter
and the pump will start dosing at top speed. Finally, just set the amount in ml (from 0 to 300).
Fig. 16- Pump Calibration Menu
Summary of Contents for DGTEC CD
Page 25: ...ADSP7000634 rev 1 3 23 09 2020 25 36 ANNEX A Overall dimensions ...
Page 27: ...ADSP7000634 rev 1 3 23 09 2020 27 36 ANNEX C Exploded views ...
Page 33: ...ADSP7000634 rev 1 3 23 09 2020 33 36 ANNEX G Menu Layouts Initial Menu User Menu ...
Page 34: ...ADSP7000634 rev 1 3 23 09 2020 34 36 Programming Menu Statistics Menu Settings Menu ...
Page 36: ...ADSP7000634 rev 1 3 23 09 2020 36 36 DGTEC CD ...