![Aqua Lung G3000SS Operation & Maintenance Manual Download Page 57](http://html1.mh-extra.com/html/aqua-lung/g3000ss/g3000ss_operation-and-maintenance-manual_2958530057.webp)
ANNUAL INSPECTION AND MAINTENANCE CHECKLIST (continued)
STEPS
PROCEDURE
INITIALS
FREE FLOW / WATER DUMP VALVE AND DEFOGGER
NOTE:
Refer to the free flow / water dump valve and defogger section of the
Operations and Maintenance Manual for guidance.
FREE FLOW / WATER
DUMP VALVE BODY
• Remove the free flow / water dump valve from
the helmet shell. Disassemble fully and visually
inspect all parts for damage. Clean as required.
WATER DUMP EXHAUST
FLAPPER VALVES
• Visually inspect for cuts and damage. Wash with
soap and water, rinse with fresh water and allow
to dry. If necessary, replace water dump / exhaust
flapper valve.
NOTE:
The flapper valves have a long
service life and are not normally replaced
annually.
O-RINGS (3)
• Replace all o-rings (3) annually.
COPPER WASHERS (2)
• Replace both copper washers annually.
VALVE SEAT
• Visually inspect and check nylon seat condition.
• If necessary, replace entire free flow / water dump
valve seat.
FREE FLOW / WATER
DUMP VALVE STEM
• Visually inspect for damage, bends and thread
condition.
• If necessary, replace with new valve stem.
57