CHAPTER 6
INSTRUMENTATION CALIBRATION
Entire Contents Copyright
2004 by Adaptive Power Systems, Inc. (APSI) • All Rights Reserved • No reproduction without written authorization from APSI.
APS-1000 Series
6 - 2
OM-001-01000-000-04.0
Calibration Setup
APS-1000 calibration is straightforward. However, you must use the appropriate
calibration instrumentation and text fixtures. In addition to test cables and connectors,
you will need:
RMS Voltmeter,
±
0.2%, at least 300 VAC
RMS Ammeter,
±
0.2%, (See Calibration Settings Table for your APS model)
Restive Load Bank, Calculated for your APS model (See following Example)
How to Calculate the Size of the Test Load (Example)
For an example, suppose your system is an APS-1010.
The High Current Calibration Procedure (later in this chapter) specifies an output
voltage of 120 VAC, for all models. From the Table of Calibration Settings, you see
the
A HI (A)
calibration test current for the (example) APS-1010P is specified at 80 A.
CALIBRATION SETTINGS
Parameter
1002
1003
1005
1010
1020
1030
1040
V LO (VAC)
150.0
150.0
150.0
150.0
150.0
150.0
150.0
V HI (VAC)
300.0
300.0
300.0
300.0
300.0
300.0
300.0
A LO (A)
3.000
3.000
30.00
30.00
30.00
30.00
30.00
A HI (A)
16.00
25.00
42.00
80.00
160.0
250.0
330.0
P LO (kW)
0.3000
0.3000
3.000
3.000
3.000
3.000
3.000
P HI (kW)
2.000
3.000
5.000
9.000
18.00
27.00
36.00
For 120 VAC and 80 A, your load should have a resistance of
The power handling capability of the test load must be at least
For 120 VAC and 80 A, your load should have a resistance of
The power handling capability of the test load must be at least