Reference Standard RS 2x30, RS 1x30
User’s Guide
RS-UG91-EN
www.appliedp.com
24
Control port description
All pins are
galvanicaly isolated (optical and relay isolation)
5.5 Serial Port (RS-232 PORT)
Setting of serial COM1 (RS-232) port can be viewed / changed by submenu System / RS-232 or by
remote interface commands:
SYSTem:SETup:COM1?
Request for COM1 setup. Answer is : baudRate(1200-115200)
,
data bits (7 or 8), parity (N=none,
O=odd, E=even) and stop bits (1 or 2) separated by commas.
SYSTem:SETup:COM1 {<baudRate>} [,{7 | 8} [,{N | O | E} [,{1 | 2}] ] ]
Setting/requesting of COM1 setup. Parameter baudRate has available values 1200, 2400, 4800, 9600,
19200, 38400, 57600 and 115200. Optional other parameters define: data bits (7 or 8), parity (N=none,
O=odd, E=even) and stop bits (1 or 2).
5.6 Configuration & Calibration settings
Configuration settings (values of parameters in all menus except Calibration menu) can be saved or
loaded by menu items Save Config and Def. Config in Main menu or by remote interface commands:
SYSTem:CONFig:SAVe
Save configuration.
SYSTem:CONFig:LOAD
Load user configuration.
SYSTem:CONFig:LOAD:DEFault
Load default configuration.
Calibration settings (values of parameters in Calibration menu) can be saved or loaded by menu
items Save Calib and Load Def. Calib in password protected Calibration menu or by commands:
SYSTem:CALib:SAVe {<password>}
Save calibration. Parameter is the password used for entering Calibration menu of device.
SYSTem:CALib:SAVe:DEFault {<AP_password>}
Save default calibration. Parameter is AP password.
SYSTem:CALib:LOAD
Load calibration.
SYSTem:CALib:LOAD:DEFault
Load default calibration.
Pin #
Name
Description
1
OUT 0
Output signals
(TTL levels)
2
OUT 1
3
OUT 2
4
OUT 3
5
SU1
Voltage dosage contacts
(max. 125V, 1A)
6
SU2
7
SI1
Current dosage contacts
(max. 125V, 1A)
8
SI2
9
IMP+
Output impulses
(RS-485 levels)
10
IMP-
11
IN0
Input signals
(TTL levels)
12
IN1
13
IN2
14
IN3
15
GND
Ground
(TTL level)