Reference Standard RS 2x30, RS 1x30
User’s Guide
RS-UG91-EN
www.appliedp.com
36
6
th
Line:
-
k
............................... meter constant of the meter under test (set in Measurement / Energy /
Impulse submenu)
7
th
Line:
-
AVG
......................... average value from all measures samples shown after getting second
sample and updated after each new sample. After finishing of test
(measuring required number of samples) is this value result of test.
8
th
Line:
-
STD
......................... standard deviation value of all measures samples shown after getting
second sample and updated after each new sample.
Middle part of window
6
th
- 8
th
Line:
-
samples
.................. list of measured samples with their sequence numbers. List can be
browsed using UP and DOWN arrows.
Right part of window
Right part of window contains list of available keys
– some of them are associated with parameters
which values are shown next to key indicator.
-
1
or
F1 / Meas
......... key for entering or leaving MEAS mode of the OPS unit. This key starts /
ends meter testing when all other parameters are set to correct values.
Actual state is indicated in end of line: ON or OFF.
-
2
or
F2
/
Calib
......... key for entering or leaving BARGRAPH / CALIB mode of the OPS unit.
Actual state is indicated in end of line: ON, SET or OFF.
-
3
or
F3
/
Input
......... key for setting input mode: LED, DISK or S0.
-
4
or
F4
/
Res.
.......... key for setting resolution: LOW = results with two decimal places and
10000 nominal impulses (100 ppm resolution) or HIGH = results with three
decimal places and 60000 nominal impulses (17 ppm resolution)
-
5
or
F5
/
Wind.
........ key for setting measurement window used by OPS during capturing marks
from rotation disks. Value can be 0-100%.
-
6
or
F6
/
Imp.
..........
key for setting of “OPS predivider” e.g. number of impulses captured by
OPS unit needed for every sample. Value can be 1-60000.
-
7
/
Samp.
................ key for setting of number of samples needed for evaluation of test. Values
can be 1-50.
-
8
/
Mode
.................. key for setting testing mode: Meter Error Test, Meter Constant Test with
value shown in [/kWh or /kVAh] or Meter Constant Test with value shown
in [/Ws or /VAs]
-
9
/
Clr
...................... key for clearing of history (list) of measured samples. Usage of this key
when testing is in progress makes reset of the test without interruption of
the test.