9.
Position the Automated Thermal Cycler (ATC) System as shown in the following image to remove
the six M3 x 10 socket head cap screws with a Hex/ Allen key.
Note:
The tightening torque is 8.00 kgfcm (+/- 10%).
10.
Separate the auto module from the base module as shown in the image to reveal the wire
connection, and unplug the front closing sensor connector, back opening sensor connector, and
the motor connector.
Chapter 4
Maintain the Automated Thermal Cycler (ATC) System
Separate the heated cover from the block
4
38
Automated Thermal Cycler (ATC) System User Guide