6.
Press
Load
to display previously protocols.
Note:
An empty protocol cannot be saved.
7.
Click
Export SiLA
to export protocol for integration with SiLA based software or scheduler.
Note:
Although SiLA exported protocol is an .xml file it should opened using a text editor in
order to see the content, otherwise they might appear empty if opened with a default browser
application. When exporting to SiLA format, no white space is allowed in stage label and protocol
file name. Only alpha numeric characters and “-“, “_” are accepted.
Chapter 2
Operate the Automated Thermal Cycler (ATC) System as a stand-alone system
Program and execute a run on the ATC
2
Automated Thermal Cycler (ATC) System User Guide
19