CSPN Capacitor Charging Unit Operation Manual
CSP-CSPN-8000/1
__________________________________________________________________________________________________________
__________________________________________________________________________________________________________
Page
14
of
63
Local / Remote
The LOCAL / REMOTE switch allows connection of remote box for operation from the laboratory or
instrument room. Using the remote box is achieved by turning the LOCAL / REMOTE switch to
remote. The high voltage can only be turned on from the instrument room and
not
from the CSP-D
Unit. The high voltage OFF button is operational from the CSP-D Unit
and
remote box. The CSP
Unit can also be keyed from the remote box via a BNC socket. The remote also monitors the
status of the key detect circuitry, and the key
LED will only illuminate on reception of a valid key pulse within the CSP-D itself, thus the
operator can also see if the interconnection cable is OK. The remote also has a FAULT LED, as
well as high voltage ON and high voltage OFF indicators.
LOCAL
Operation indicated by LED on – Switch OUT.
REMOTE
Operation indicated by LED out – Switch IN