Set the sampling interval in minutes or seconds. The sampling
interval is how often a measurement is taken and logging interval is
how often the data is saved. The logged data is the average of the
samples. The logging interval must be evenly divided by the
sampling interval. For example if the logging interval is 5 minutes
and the sampling interval is 2 minutes it causes an error. But a
sampling interval of 1 minute is acceptable.
Before clicking ‘Get Logged Data’ it is important to set the time of
the last logged data point. This is used to back calculate the
timestamps for the remaining data points. If you just unplugged
the sensor and plugged it into the computer the preloaded day
and time should be sufficient.
Click ‘
Get Logged Data
’ to save the data to your computer. You
will be asked where you want to save the data.
Click ‘
Erase Data
’ to erase all the save data. This can’t be undone.
To use additional SP-422 devices, open additional ApogeeConnect software windows. The device serial number will display in the lower
left hand corner of the corresponding software window. Devices may be selected by serial number in the tool bar.